Related Experiment Video
Updated: May 2, 2026

Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Lumbar fusion using a carbon-fiber PEEK pedicle screw system combined with a carbon-fiber PEEK TLIF cage: a
Marcus Eif1, Thomas Forster2, Christoph Fleege3
1Städtisches Klinikum Görlitz gGmbH, Klinik für Neurochirurgie, Girbigsdorfer Straße 1-3, 02828, Görlitz, Germany.
Background:
Carbon-fiber-reinforced polyether ether ketone (CF-PEEK) is a radiolucent, non-metallic implant material used for instrumented lumbar spondylodesis. Clinical studies of pedicle screw systems employing this material, especially for degenerative indications, are scant.
Methods:
We conducted a multicenter, prospective clinical study to assess clinical and radiographic outcomes in patients with symptomatic degenerative lumbar disk disease, including degenerative spondylolisthesis treated with a CF-PEEK pedicle screw and a transforaminal lumbar interbody fusion (TLIF) cage system. We followed up the participants for two years postoperatively to collect clinical data (via the Oswestry Disability Index, Core Outcome Measures Index, and Visual Analog Scale), radiographic parameters (functional X-rays) to assess fusion status, and any complications.
Results:
In total, 86 patients were recruited. During the study, 21 patients (24.4%) dropped out, including 5 (5.8%) who underwent explantation of the study device(s). At the final follow-up, the fusion rate was 98.6% (95% confidence interval, 92.7-100.0%). All clinical parameters improved significantly. There were no complications potentially attributed to the implant material.
Conclusions:
The results demonstrate a fusion rate similar to that of metallic implant systems with the use of a CF-PEEK pedicle screw and a TLIF cage system. Further studies with larger samples are needed to substantiate this finding.
Trial Registration:
The study was registered at ClinicalTrials.gov (NCT02087267). Date of registration: March 12, 2014.