Related Experiment Video
Updated: May 23, 2025

Author Spotlight: Learning Systematic Bronchoscopy in a Simulation-Base Setting
Published on: June 23, 2023
A Micro-costing Analysis of Single-use and Reusable Flexible Bronchoscope Usage in the Bronchoscopy Service at A
Javier Flandes1, Andrés Giménez1, Susana Álvarez1
1Department of Respiratory, Bronchoscopy and Interventional Pumonology Service, Fundación Jiménez-Díaz University Hospital, Madrid, Spain.
Background:
Single-use flexible bronchoscopes (SFBs) are increasingly used to minimize cross-infection risk, particularly in immunocompromised and intensive care unit patients. However, broader adoption requires cost analysis. We conducted a 1-year cost-minimization analysis comparing SFBs and reusable flexible bronchoscopes (RFBs) at a tertiary care university hospital.
Methods:
We evaluated the costs per procedure, considering capital equipment, maintenance, repair, reprocessing, and overhead costs. We also analyzed the impact of annual procedure volume on costs and performed a sensitivity analysis to assess the effect of uncertainty on costs.
Results:
A total of 1394 bronchoscopies were performed. RFBs were less expensive for an annual volume of >50 bronchoscopies/year, with a 22% lower cost per procedure than that for SFBs (€203 vs. €259). This cost advantage became increasingly favorable with an increasing number of procedures, reaching a plateau after exceeding 250 bronchoscopies/year. The capital equipment, the annual number of bronchoscopies, and reprocessing were the major cost drivers for RFBs. During nonworking hours, the cost per procedure of RFBs ranged from €349.45 to €392.29. Using RFBs during interventions involving a high risk of bronchoscope damage (frequency of damage >10%) would increase the cost per bronchoscopy to >€263 (exceeding the cost of SFBs).
Conclusion:
RFBs were 22% less expensive than SFBs for services with a moderate to high volume of bronchoscopies. However, this difference could not justify using RFBs in patients with a high cross-infection risk. SFBs might be less costly for procedures outside working hours and interventions involving a high risk of bronchoscope damage.
Related Concept Videos
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...