A patient-matched prosthesis for thumb amputations: Design, mechanical and functional evaluation
Federico Stacchiotti1, Chiara Bregoli2, Rubens Ferrari1
1Istituto di BioRobotica, Scuola Superiore Sant'Anna, Viale Rinaldo Piaggio 34, 56025, Pontedera, Pisa, Italy.
Medical Engineering & Physics
|March 8, 2025
Summary
This study introduces the first patient-matched osseointegrated prosthesis for thumb amputations, restoring daily hand function. The novel device demonstrates mechanical performance suitable for daily activities, with potential failure points identified externally.
Area of Science:
- Biomedical Engineering
- Reconstructive Surgery
- Prosthetics
Background:
- Thumb amputations significantly impair hand functionality, with existing treatments like microsurgery and vacuum prostheses having limitations.
- Osseointegration offers a promising alternative, but patient-matched devices for thumb amputations are scarce due to variations in residual bone anatomy.
- Current prosthetic solutions lack customization, failing to address the diverse characteristics of remaining bone in amputated limbs.
Purpose of the Study:
- To present the first patient-matched osseointegrated medical device specifically designed for thumb amputations.
- To optimize the design of the osseointegrated fixture, abutment, and external prosthesis through computational and experimental analysis.
- To evaluate the mechanical performance of the novel device for restoring thumb functionality in daily activities.
Main Methods:
- Development of a three-component system: an implanted osseointegrated fixture, an abutment, and an external digital prosthesis.
- Design optimization using computational and experimental analyses focusing on preserving the implant fixture and bone integrity.
- Mechanical testing, including pinch force analysis, comparing device performance against healthy subject data (approx. 80 N maximum pinch force).
Main Results:
- The novel patient-matched osseointegrated device successfully achieves mechanical performance required for daily activities.
- Computational and experimental analyses confirmed the device's ability to meet functional demands.
- Failure analysis indicated that under loads exceeding daily use, failure is likely to occur in the external abutment component, not the internal fixture or bone.
Conclusions:
- The developed patient-matched osseointegrated prosthesis represents a significant advancement in treating thumb amputations.
- The device shows potential for restoring crucial hand functionality, with a safety mechanism where excessive force leads to external component failure.
- Further clinical studies are necessary to validate the device's long-term efficacy and gather patient/surgeon feedback, particularly regarding the bone-implant interface dynamics.