Visual acuities as measured with an automatic intelligent visual acuity chart projector and standard logarithmic
Mingyue Luo1,2,3, Xinyu Liu1,2,3, Dingding Zhang4
1Department of Ophthalmology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Frontiers in Medicine
|March 11, 2025
Summary
The novel intelligent projector visual acuity (IP) chart shows excellent repeatability and consistency with standard charts for visual acuity (VA) measurement. This new IP chart offers reliable VA testing comparable to traditional methods.
Area of Science:
- Ophthalmology and Vision Science
- Medical Devices and Technology
Background:
- Accurate visual acuity (VA) measurement is crucial for diagnosing and monitoring eye conditions.
- Traditional visual acuity charts may have limitations in modern clinical settings.
- Novel technologies like intelligent projector (IP) charts aim to improve VA testing efficiency and accuracy.
Purpose of the Study:
- To evaluate the repeatability of VA measurements using a new intelligent projector (IP) chart (LSJ-IVAC-6000A).
- To assess the consistency of VA measurements between the IP chart and the standard logarithmic visual acuity (SL) chart (GB11533-2011).
Main Methods:
- Prospective comparative study involving 53 subjects for repeatability and 100 subjects for consistency analysis.
- Intraclass correlation coefficient (ICC) and Bland-Altman analyses were used to assess repeatability and consistency.
- Visual acuity was recorded using the logMAR method, and test duration was timed.
Main Results:
- The IP chart demonstrated high repeatability with ICC values of 0.968 (right eye) and 0.964 (left eye).
- Good consistency was observed between the IP and SL charts, with ICC values of 0.946 (right eye) and 0.817 (left eye).
- The IP chart showed comparable test times to the SL chart, with no significant difference (p=0.668).
Conclusions:
- The novel intelligent projector visual acuity chart exhibits excellent repeatability.
- The IP chart demonstrates overall good consistency with the standard logarithmic visual acuity chart.
- The IP chart is a reliable tool for visual acuity measurement in clinical practice.
Keywords:
consistency analysisintelligent projector visual acuity chartrepeatability analysisstandard logarithmic visual acuity chartvisual acuity