Related Experiment Video
Updated: Jul 11, 2026

A Study of the Complexation of Mercury(II) with Dicysteinyl Tetrapeptides by Electrospray Ionization Mass Spectrometry
Published on: January 8, 2016
Neutron activation determination of mercury in waters after preconcentration by flotation of dithizone-mercury
Abstract:
Mercury in water samples, at levels below parts per billion, was collected by dithizone flotation. The statistical detection limit of Hg in seawater was 4 times better than that with a hydrous iron oxide--APDC flotation system in neutron activation determination. The same floated dithizone precipitate can be used repeatedly for collection from several volumes of the water sample. The Hg content in seawater was 0.017 micrograms/L. Average recovery from waters containing 0.025 micrograms/L, 0.05 micrograms/L and 0.1 micrograms/L was 98%. The method was applied successfully to the determination of Hg in NBS 1641b. Methyl mercury is also collected by the procedure described.
Related Concept Videos
Voltammetric Techniques: Linear-Scan (E vs Time)
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...
Washing, Drying, and Ignition of Precipitates
Extraction: Advanced Methods