Decrease in the internal quality control intermediate reproducibility imprecision of Cystatin C results in China in
Zhixin Zhang1, Jie Zeng1, Wei Wang1
1National Center for Clinical Laboratories/Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing, China.
Annals of Clinical Biochemistry
|March 14, 2025
Summary
Internal quality control data show that cystatin C testing imprecision has improved in China from 2014 to 2023. While overall performance is better, biological variation-based targets remain challenging for some labs.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Quality Control
Background:
- Internal quality control (IQC) data are crucial for assessing laboratory testing performance.
- Cystatin C is a biomarker used for estimating glomerular filtration rate.
- Evaluating intermediate reproducibility imprecision is essential for reliable diagnostic results.
Purpose of the Study:
- To evaluate the intermediate reproducibility imprecision of cystatin C results in Chinese laboratories using IQC data from 2014 to 2023.
- To assess laboratory performance against various established quality specifications.
- To identify trends and differences in imprecision based on reagent manufacturers and analytical platforms.
Main Methods:
- Analysis of IQC data for cystatin C from 255 to 1814 laboratories annually (2014-2023).
- Calculation of coefficient of variation (CV) to measure laboratory imprecision.
- Assessment of pass rates against five performance specifications, including 1/3 total allowable error (TEa) and biological variation (BV).
- Stratified analysis by QC material concentration, reagent manufacturer, and instrument platform.
Main Results:
- Significant decrease in cystatin C imprecision (CV) from 5.1% in 2014 to 3.3% in 2023.
- Upward trend in pass rates based on 1/3 TEa, reaching 88% in 2023.
- Pass rates for BV-based specifications remained below 80% for many laboratories.
- Roche Diagnostics showed low imprecision; BSBE-Abbott Architect achieved 100% pass rate for 1/3 TEa in 2023.
Conclusions:
- Continuous improvement in cystatin C intermediate reproducibility imprecision in China.
- Current BV-based performance specifications for cystatin C may not be universally achievable in Chinese laboratories.
- Significant differences in imprecision exist across various analytical systems, warranting further attention.