Validating image-derived input functions of dynamic 18F-FDG long axial field-of-view PET/CT studies
Charlotte L C Smith1,2, Gerben J C Zwezerijnen1,2, Marijke E den Hollander1
1Department of Radiology and Nuclear Medicine, Amsterdam UMC Location Vrije Universiteit Amsterdam, Amsterdam, Netherlands.
Aim/Background:
Dynamic PET imaging requires an input function typically obtained through blood sampling. Image-derived input functions (IDIFs) of the ascending aorta (AA), aortic arch, descending aorta (DA), or left ventricle (LV) offer non-invasive alternatives, especially with long-axial field-of-view (LAFOV) PET/CT systems enabling whole-body dynamic 1⁸F-FDG imaging. This study aimed to validate uncorrected IDIFs derived from the AA, DA, aortic arch, and LV by comparing them to (late) venous whole-blood in patients undergoing LAFOV PET/CT.
Methods:
Eleven oncology patients who underwent 70-min dynamic 18F-FDG PET/CT scans on a LAFOV PET/CT system after receiving an intravenous bolus injection of 3.0 MBq/kg were included. Seven venous blood samples were collected manually at approximately 5, 10, 15, 25, 35, 45, and 60 min post-injection (pi) and compared to IDIFs derived from the AA, aortic arch, DA, and LV. Bias between IDIFs and venous blood samples was assessed at each time point.
Results:
IDIF accuracy relative to venous blood samples improved over time, with a median percentage bias <10% after 25 min pi. At 60 min pi, the aortic arch showed the smallest bias (median -1.1%, IQR 5.9%), followed by the AA (2.5%, IQR 7.0%), DA (5.1%, IQR 8.6%), and LV (7.4%, IQR 7.6%).
Conclusion:
The high precision of aorta-derived IDIFs suggests that IDIFs are a reliable alternative to manual blood sampling for dynamic 18F-FDG PET imaging on a LAFOV PET/CT system. Using IDIFs reduces variability, simplifies protocols, minimizes radiation exposure, and enhances patient safety with a non-invasive approach.
More Related Videos
Related Concept Videos
BIBO stability of continuous and discrete -time systems
To determine the BIBO stability, the convolution integral is utilized when a bounded continuous-time input is applied to a Linear Time-Invariant (LTI) system....
First Order Systems
When a first-order system is subjected to a unit-step input, its response is characterized by its transfer function. By applying the Laplace transform of the unit-step input to the transfer function, expanding the...
Convolution Properties II
The width property indicates that if the durations of input signals are T1 and T2, then the width of the output response equals the sum of both durations, irrespective of the shapes of the two functions. For instance, convolving two rectangular pulses with durations of 2 seconds and 1 second results in a function with a width of 3 seconds.
The area property asserts that the area under the...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Multi-input and Multi-variable systems
In the absence...
Singularity Functions for Shear