Related Experiment Video
Updated: May 22, 2025

Making MR Imaging Child's Play - Pediatric Neuroimaging Protocol, Guidelines and Procedure
Published on: July 30, 2009
Optimizing T2* imaging for adolescent and young adult patients at 7 T
Sergio Valencia1,2, Fedel Machado-Rivas3,4, Arman Avesta3,4
1Massachusetts General Hospital, Boston, USA. svalenciavasquez@mgh.harvard.edu.
Background:
T2*-weighted imaging at 7 T offers detailed visualization of brain structures, but image quality and artifacts depend on echo time (TE) adjustments. Optimizing TE is crucial for tissue contrast and artifact minimization.
Objective:
To evaluate the impact of TE on tissue contrast and image quality in T2*-weighted sequences at 7 T in adolescents and young adults.
Materials And Methods:
Ten adolescent and young adult patients underwent 7-T MRI with multi-echo T2*-weighted sequences. Six TEs (8.1 ms to 36 ms) were acquired. Signal contrast ratios (CR) for seven brain regions-caudate nuclei, corpus callosum genu, frontal cortex, cortical veins, globi pallidi, medullary veins, and left optic radiation-were analyzed. Two blinded neuroradiologists assessed image quality and artifact severity using a 4-point Likert scale (IQS). Statistical trends were analyzed using the Jonckheere-Terpstra test. A P-value < 0.05 was considered statistically significant.
Results:
The study cohort consisted of 4 male and 6 females; the median age of the patients was 16 years (range 15-23 years). CR increased significantly with higher TEs for most regions except the caudate, where CR decreased (P < 0.05). Longer TEs led to greater artifact severity in the brainstem, temporal, occipital, and frontal lobes (P < 0.02), but not in parietal lobes (P > 0.05). Kappa agreement for IQS was 0.76.
Conclusion:
TE significantly affects contrast and artifacts in 7-T T2*-weighted imaging. TEs between 20 and 30 ms offer the best balance between tissue contrast and artifact severity, optimizing image quality for clinical and research applications.
Related Concept Videos
Magnetic Resonance Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...