Related Experiment Video
Updated: May 21, 2025

Endotracheal Intubation in Mice via Direct Laryngoscopy Using an Otoscope
Published on: April 5, 2014
Comparison of Macintosh Direct Laryngoscope with the C-MAC and Tuoren Videolaryngoscopes in Facilitating Endotracheal
Rajender Kumar1, Rakesh Kumar2, Naveen Kumar1
1Department of Critical Care, Dr Baba Saheb Ambedkar Medical College and Hospital, Rohini, New Delhi, India.
Background:
Videolaryngoscopes (VLS) frequently provide superior views of the glottis compared with traditional direct laryngoscopy (DL), especially during unexpected difficult airways. Chest compressions during attempts to intubate the trachea make it a difficult situation. Although VLS have been compared with DL for intubation during resuscitation, there is a paucity of literature comparing VLS with integrated screen and distant screen with DL during continued manual chest compressions.
Materials And Methods:
This was a prospective, randomized, crossover observational manikin study. A total of 20 participants performed intubation, while manual chest compression was continuing, with each of the three devices on six occasions, but in different, randomized order. The primary outcome parameter was the total time taken for successful intubation. The secondary outcome criteria included the number of attempts, ease of intubation, and the device preference.
Results:
Time taken for successful intubation and ease of intubation were significantly better with C-MAC VLS and DL as compared with Tuoren VLS (C-Mac vs Tuoren p < 0.000 for both; DL vs Tuoren p < 0.001 for time and p = 0.021 for ease). There was no significant difference between C-MAC and DL (p = 1.0 for time and p = 0.69 for ease). There was no significant difference with regard to the number of attempts for successful intubation with any of these devices (p = 0.310). C-Mac was the most and Tuoren was the least preferred device.
Conclusion:
C-MAC VLS and Macintosh DL are significantly easier to use and require significantly less time to achieve successful intubation as compared with the Tuoren VLS during continued manual chest compression. Among the devices tested, C-MAC VLS was the most preferred for endotracheal intubation during uninterrupted manual chest compressions.
How To Cite This Article:
Kumar R, Kumar R, Kumar N. Comparison of Macintosh Direct Laryngoscope with the C-MAC and Tuoren Videolaryngoscopes in Facilitating Endotracheal Intubation during Uninterrupted Manual Chest Compression: A Randomized Crossover Manikin Study. Indian J Crit Care Med 2025;29(2):113-116.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...