Effect of different airway devices on ventilation during cardiopulmonary resuscitation
N Segond1, M Fischer2, J Fontecave-Jallon3
1Emergency Department and Mobile Intensive Care Unit, University Hospital of Grenoble Alpes, Grenoble, France; Univ. Grenoble Alpes, CNRS, UMR 5525, VetAgro Sup, Grenoble INP, TIMC, 38000 Grenoble, France.
Purpose:
This study compared face mask, supraglottic airway device (SGA), and endotracheal tube (ETT) ventilation with mechanical ventilation (MV) during cardiopulmonary resuscitation (CPR) in the flat position and with head and thorax elevation (HTE).
Methods:
Using thawed, fresh-frozen human cadavers this randomized cross-over study compared face mask, SGA, and ETT ventilation using an automated ventilator in the flat and HTE positions. Tidal volume (TV) was set to 8 mL/kg ideal predicted body weight, and expiratory TV (VTe) (mL/kg) was the primary endpoint. Secondary endpoints included inspiratory TV (VTi), maximal inspiratory airway pressure (Pmax), and leakage between inspiratory and expiratory tidal volumes (VTi-VTe).
Results:
Data from 8 cadavers and 2302 ventilation cycles were analyzed. In the flat position, VTe was 7.66 ± 3.75 with ETT, 5.01 ± 3.14 with SGA, and 5.63 ± 2.83 with face mask, respectively. A mixed linear model showed the airway device significantly impacted VTe, VTi, Pmax, and VTi-VTe (p < 0.001 for each). Compared with intubation, both face mask and SGA ventilation resulted in lower VTe, lower Pmax, and higher VTi-VTe (p < 0.001 for each). No significant differences were observed between face mask and SGA ventilation. There were higher VTe and lower VTi-VTe values (p < 0.001 for each) with HTE versus the flat position (p < 0.001).
Conclusions:
In human cadavers undergoing CPR, mechanical ventilation through a face mask or SGA versus an ETT was associated with lower VTe, lower Pmax, and higher leakage values in human cadavers during CPR. Head and thorax elevation reduce face mask and SGA airway leakage during CPR and increase VTe.
More Related Videos
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...