Related Experiment Video
Updated: May 20, 2025

Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
Shielding resources for four common radiopharmaceuticals utilized for imaging and therapy: Tc-99m, F-18, I-131, and
Michael Oumano1,2,3, Richard Wendt4, James Botti5
1Department of Medicine and Biological Sciences, Brown University, Providence, Rhode Island, USA.
Introduction:
The use of radioactive materials in the United States has been tightly regulated by the Nuclear Regulatory Commission and other entities for many decades. In 2015, however, the Joint Commission began to require hospital-based nuclear medicine departments to conduct shielding designs and evaluations for radioactive material areas, mirroring established x-ray practices. NCRP Report No. 147 guides diagnostic medical x-ray shielding, but obviously cannot be used alone for nuclear medicine applications. The rising demand for theranostic nuclear medicine shielding evaluations particularly necessitates updated focused guidance, the aim of this report.
Methods:
Monte Carlo simulations were conducted using GATE software to analyze the effects of various barriers on the transmission of radioactive emissions. The simulations used point sources of Tc-99m, F-18, I-131, and Lu-177 and evaluated dose deposition to blocks of tissue using Dose Actors. Different ceiling heights (ranging from 10-16 feet) were also tested for scattering effects. The Archer equation was employed to fit transmission curves and estimate required barrier thicknesses.
Results:
Broad beam transmission factors and Archer fitting parameters are reported for various materials including lead, gypsum, concrete (light weight and normal weight), glass, and steel. A sample shielding calculation is provided for a wall separating Lu-177 dotatate patients from an adjacent office to maintain public dose limits. Relevant occupancy factors are also provided.
Conclusions:
While Lu-177 has a relatively low exposure rate constant, shielding may be necessary for high-volume therapies like Lu-177 DOTATATE and Lu-177 vipivotide tetraxetan PSMA. Shielding involves accounting for broad radiation beams and requires thorough characterization of radiation buildup. When shielding to the typical height of 7 feet, scatter from ceilings and floors is negligible for transmission above 10%, but severely limits the ability to shield for transmission below 1%.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:11Y-90 Radioembolization and PD-1 Inhibitor as Neoadjuvant Treatment in Hepatocellular Carcinoma
Published on: May 24, 2024
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...