Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Spectrophotometry: Introduction01:16

Spectrophotometry: Introduction

Spectrophotometry is the quantitative measurement of the absorption, reflection, diffraction, or transmission of electromagnetic radiation through a material as a function of the intensity and wavelength of the radiation. A spectrophotometer is a device used to measure the change in the radiation intensity caused by its interaction with the material.
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
UV–Vis Spectroscopy: Beer–Lambert Law01:09

UV–Vis Spectroscopy: Beer–Lambert Law

The Beer-Lambert law describes the relationship between absorbance and concentration, which combines the principles established by scientists Johann Heinrich Lambert and August Beer. Lambert's law states that when light passes through a medium, the loss in intensity is directly proportional to the original intensity and the path length of the light. Beer's law proposed that the transmittance of a solution remains constant if the product of concentration and path length is constant. The modern...
UV–Vis Spectrometers01:14

UV–Vis Spectrometers

The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell. Samples for...
Atomic Emission Spectroscopy: Lab01:29

Atomic Emission Spectroscopy: Lab

AES is a powerful analytical technique, especially effective when used with plasma sources, producing abundant spectra in characteristic emission lines. The Inductively Coupled Plasma (ICP), in particular, yields superior quantitative analytical data due to its high stability, low noise, low background, and minimal interferences under optimal experimental conditions. However, newer air-operated microwave sources are emerging as promising alternatives that could be more cost-effective than...
Flame Photometry: Overview01:02

Flame Photometry: Overview

Flame photometry, also known as flame emission spectrometry, is a technique used for the qualitative and quantitative analysis of elements present in a sample using a flame as the source of excitation energy. The concept of flame photometry was realized in the early 1860s by Kirchhoff and Bunsen, who discovered that specific elements emit characteristic radiation when excited in flames. The first instrument developed for this purpose was used to measure sodium (Na) in plant ash using a Bunsen...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Can processed images be used to determine the modulation transfer function and detective quantum efficiency?

Journal of medical imaging (Bellingham, Wash.)·2024
Same author

Reduced Total Airway Count and Airway Wall Tapering after Three-Years in Ex-Smokers.

COPD·2023
Same author

Theoretical comparison of energy-resolved and digital-subtraction angiography.

Medical physics·2022
Same author

Frequency-dependent signal and noise in spectroscopic x-ray imaging.

Medical physics·2020
Same author

Cascaded systems analysis of charge sharing in cadmium telluride photon-counting x-ray detectors.

Medical physics·2018
Same author

Report of AAPM Task Group 162: Software for planar image quality metrology.

Medical physics·2017
Same journal

Correction to "On the shape of the radiation survival curve in tumor spheroids: The role of oxygen heterogeneity".

Medical physics·2026
Same journal

Multi-view constrained semi-supervised vertebra detection for 3D ultrasound spine volume.

Medical physics·2026
Same journal

Accuracy of quantitative <sup>177</sup>Lu SPECT/CT imaging: A systematic review.

Medical physics·2026
Same journal

Physics-constrained dual-domain network for CBCT reconstruction from orthogonal X-rays in gynecologic radiotherapy.

Medical physics·2026
Same journal

Decomposition-based harmonization for quantitative PET imaging across scanners and radiotracers.

Medical physics·2026
Same journal

Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.

Medical physics·2026
See all related articles

Related Experiment Video

Updated: Jun 16, 2026

Optical Scatter Microscopy Based on Two-Dimensional Gabor Filters
14:58

Optical Scatter Microscopy Based on Two-Dimensional Gabor Filters

Published on: June 2, 2010

9.5K

Robust Fourier-based slanted-edge method to measure scatter ratio.

Lisa M Garland1, Ian A Cunningham1

  • 1Imaging Research Laboratories, Robarts Research Institute, and Department of Medical Biophysics, Western University, London, Ontario, Canada.

Medical Physics
|March 25, 2025
PubMed
Summary
This summary is machine-generated.

A new method accurately measures the scatter-to-primary ratio (SPR) in X-ray imaging. This technique, utilizing Fourier analysis of edge profiles, works across various radiographic and mammographic conditions, improving image quality.

Keywords:
Fourier domainX‐ray imagingX‐ray scatter

More Related Videos

Measuring Spatially- and Directionally-varying Light Scattering from Biological Material
11:57

Measuring Spatially- and Directionally-varying Light Scattering from Biological Material

Published on: May 20, 2013

12.6K
Measuring the Behavioral Effects of Intraocular Scatter
05:10

Measuring the Behavioral Effects of Intraocular Scatter

Published on: February 18, 2021

3.3K

Related Experiment Videos

Last Updated: Jun 16, 2026

Optical Scatter Microscopy Based on Two-Dimensional Gabor Filters
14:58

Optical Scatter Microscopy Based on Two-Dimensional Gabor Filters

Published on: June 2, 2010

9.5K
Measuring Spatially- and Directionally-varying Light Scattering from Biological Material
11:57

Measuring Spatially- and Directionally-varying Light Scattering from Biological Material

Published on: May 20, 2013

12.6K
Measuring the Behavioral Effects of Intraocular Scatter
05:10

Measuring the Behavioral Effects of Intraocular Scatter

Published on: February 18, 2021

3.3K

Area of Science:

  • Medical Physics
  • Radiological Imaging
  • Image Quality Assessment

Background:

  • X-ray scatter degrades image contrast and adds noise, impacting diagnostic accuracy.
  • Minimizing scatter is crucial for advanced techniques like dual-energy and energy-subtraction radiography.
  • Current scatter measurement methods are often impractical for routine radiography.

Purpose of the Study:

  • To develop a versatile method for measuring the scatter-to-primary ratio (SPR).
  • To enable SPR measurement across diverse radiographic and mammographic scenarios, including conditions with and without scatter equilibrium.
  • To provide a practical tool for optimizing X-ray imaging parameters.

Main Methods:

  • Utilized Fourier theory to analyze the low-frequency drop (LFD) of the Fourier transform of a normalized edge profile derivative.
  • Developed a normalization technique to eliminate the requirement for scatter equilibrium.
  • Validated the method through both experimental measurements and computer simulations.

Main Results:

  • The novel method successfully measured SPR without requiring scatter equilibrium.
  • The technique inherently accounts for primary beam non-uniformities, such as divergence and Heel effect.
  • Simulations confirmed the method's robustness across various scatter and beam conditions; experimental results aligned with existing methods under scatter equilibrium.

Conclusions:

  • A straightforward and effective method for measuring SPR in radiography and mammography has been established.
  • The technique, based on Fourier LFD from normalized slanted-edge profiles, is validated experimentally and theoretically.
  • This method is applicable to a broad spectrum of clinical X-ray imaging situations.