Effect of Plasma Air Purifiers on Infection Rates in Orthopedic Surgery

Anders Persson1,2, Isam Atroshi3,4, Thomas Tyszkiewicz5

  • 1Division of Orthopaedics, Department of Clinical Sciences, Karolinska Institutet, Danderyd Hospital, Sweden.

NEJM Evidence
|March 25, 2025
PubMed
Abstract

Insights

Low-cost air purifiers did not reduce surgical site infections (SSI) in orthopedic surgery patients. This study found no significant difference in infection rates between operating rooms with active or inactive air purifiers.

Area of Science:

  • Medical research
  • Infectious disease control
  • Public health

Background:

  • Surgical site infections (SSI) are a significant concern in orthopedic surgery.
  • Airborne transmission is a potential cause of SSI.
  • Standard ventilation systems have limited efficacy in reducing SSI rates.

Purpose of the Study:

  • To investigate the effectiveness of low-cost air purifiers in reducing SSI rates after orthopedic surgery.
  • To evaluate the impact of active air purifiers on infection markers in operating rooms.

Main Methods:

  • A nationwide, multicenter, double-blind, cluster-randomized crossover, placebo-controlled superiority trial.
  • Inclusion of all patients undergoing orthopedic surgery.
  • Comparison of surgery in operating rooms with active vs. inactive air purifiers (Novaerus NV800).
  • Primary endpoint: SSI within 12 weeks, defined by a postsurgery infection marker (PSIM).

Main Results:

  • Analysis of 40,547 patients (19,869 intervention, 20,678 control).
  • PSIM rate was 9.2% in the intervention group and 9.4% in the control group (OR 0.98, 95% CI 0.91-1.05).
  • No significant difference in SSI rates observed across subgroups.

Conclusions:

  • Wall-mounted plasma air purifiers did not reduce the PSIM rate in orthopedic surgery.
  • The addition of these air purifiers is not effective in modern operating rooms with standard ventilation.