The random-zero versus the standard mercury sphygmomanometer: a systematic blood pressure difference
American Journal of Epidemiology
|February 1, 1985
Summary
Random-zero and standard mercury sphygmomanometers show systematic differences in blood pressure readings. These devices should not be used interchangeably in epidemiologic studies due to measurement variations.
Area of Science:
- Cardiovascular research
- Medical instrumentation
Background:
- Sphygmomanometers are crucial for blood pressure measurement in epidemiological studies.
- Frequent interchangeable use of random-zero and standard mercury devices raises concerns about data consistency.
Purpose of the Study:
- To investigate systematic differences between random-zero and standard mercury sphygmomanometers.
- To determine if these devices can be used interchangeably in epidemiological research.
Main Methods:
- Simultaneous measurement of systolic and diastolic blood pressures using both sphygmomanometers.
- Controlled experiments with calibrated devices, double stethoscopes, and blinded technicians.
- Analysis of readings at a deflation rate of 2 mmHg/second, with adjustments for confounding factors.
Main Results:
- Random-zero sphygmomanometers yielded systematically lower readings: -0.9 mmHg (systolic), -1.8 mmHg (diastolic phase IV), -1.8 mmHg (diastolic phase V).
- Differences remained significant (p < 0.001) after adjusting for age, sex, heart rate, blood pressure, observer, and temperature.
- Eliminating residual mercury from random-zero devices reduced the observed measurement discrepancies.
Conclusions:
- A systematic difference exists between random-zero and standard mercury sphygmomanometers.
- The discrepancy is likely due to the mercury column height in the random-zero device.
- Interchangeable use of these sphygmomanometers in epidemiological studies is not recommended.