Related Experiment Video
Updated: Jun 20, 2026

Surgical Technique for the Implantation of Tissue Engineered Vascular Grafts and Subsequent In Vivo Monitoring
Published on: April 3, 2015
Six-Month Patency of Long Carotid Bypass Grafts Constructed with In-Body Tissue Architecture-Induced Small-Diameter
Kazuki Mori1, Tadashi Umeno1, Takayuki Kawashima1
1Department of Cardiovascular Surgery, Oita University, Yufu 879-5593, Japan.
Abstract:
This study investigated the long-term patency of regenerative Biotube grafts and discusses their feasibility as an alternative to autologous vein grafts for peripheral artery disease. Six Biotubes with a diameter of 4 mm were autologously fabricated in recipients using in vivo tissue engineering (in-body tissue architecture) technology and implanted as carotid artery bypass grafts in a goat model. All six grafts remained patent at 6 months despite exceeding 10 cm in length, demonstrating their biocompatibility and durability. Histological analysis revealed neointima formation, endothelialization, and minimal inflammation. However, in one goat, a graft developed stenosis, while another showed dilatation. These findings demonstrate the use of Biotubes as a viable option for peripheral vascular reconstruction as tissue-engineered vascular grafts. However, further optimization is needed to address emerging issues with their use, such as stenosis and aneurysm formation, to improve long-term patency.