Background:
The best management of irreparable rotator cuff tears remains uncertain, with multiple new techniques introduced over the past 2 decades. Two options for treatment are arthroscopic debridement and biceps tenotomy, or the subacromial balloon spacer. Early trial results favored the former option, but the 2-year results have not yet been reported.
Purpose:
To report the 2-year follow-up outcomes of the START:REACTS trial, investigating the use of a subacromial balloon spacer for irreparable rotator cuff tears of the shoulder.
Study Design:
Randomized controlled trial; Level of evidence, 1.
Methods:
Eligible participants had an irreparable rotator cuff tear, intrusive symptoms requiring surgery, and previous unsuccessful nonoperative care. Participants were randomized 1:1 to debridement of the subacromial space with biceps tenotomy (debridement only) or the same procedure with the addition of the subacromial balloon spacer (debridement with device). The 12-month primary outcome was previously reported; this article presents the 24-month results. Linear regression models were used to analyze the 24-month data.
Results:
Recruitment stopped early at the preplanned interim analysis, with 117 participants in the trial. A total of 99 (85%) participants out of 117 were followed up to 24 months. At 24 months, a significant difference in the Oxford Shoulder Score was not found (95% CI, -7.9 to 0.4; P = .08). The Western Ontario Rotator Cuff score (mean difference, -10.1; 95% CI, -19.5 to -0.8; P = .04) and Patient Global Impression of Change (odds ratio, 0.4; 95% CI, 0.2 to 0.8; P = .015) were found to significantly favor debridement only. The EQ-5D-5L (mean difference, -0.009; 95% CI, -0.107 to 0.088; P = .85) and satisfaction scores (odds ratio, 0.6; 95% CI, 0.3 to -1.2; P = .14) were not significantly different. Complications were evenly matched between groups over 24 months.
Conclusion:
Participants continued to show better results in the debridement-only group compared with the group who had debridement with the InSpace balloon. Therefore, we do not recommend the subacromial balloon spacer for the treatment of irreparable rotator cuff tears.