A concept of volume wireless receive-only coil for 1.5T MRI
Aleksandr Fedotov1, Pavel Tikhonov1, Viktor Puchnin1
1School of Physics and Engineering, ITMO University, St. Petersburg, Russian Federation.
Abstract:
Wireless radio frequency coils offer an alternative to conventional cable-connected coils due to their compatibility with multiple vendor MRI systems and reduced electromagnetic interaction with the environment of the MRI scanner. However, wireless coils being inductively coupled with a transceiver body coil require manual input power calibration due to the significant increase of a body coil transmit efficiency locally in the region of interest and disturbance of B1+ homogeneity complicating routine scanning procedures. This study aims to implement the concept of a wireless receive-only coil for female breast MRI at 1.5T. The approach combines the advantages of wireless coils to increase signal to noise ratio of transceiver body coil in the target region of interest and the ability to perform the automatic reference voltage calibration.