3D Ultrasound Imaging: Fast and Cost-effective Morphometry of Musculoskeletal Tissue
Published on: November 27, 2017
Sonographic image texture features in muscle tissue-mimicking material reduce variability introduced by probe angle
Dustin J Oranchuk1, Katie L Boncella2, Daniella Gonzalez-Rivera3
1Muscle Morphology, Mechanics, and Performance Laboratory, School of Medicine, University of Colorado Anschutz Medical Campus, Aurora, Colorado, United States; Department of Physical Medicine and Rehabilitation, University of Colorado, Anschutz Medical Campus, Aurora, Colorado. dustin.oranchuk@cuanschutz.edu.
Abstract:
Cost-effective and portable ultrasonography offers a promising approach for monitoring skeletal muscle damage and quality in many contexts. However, echogenicity analysis relies on precise transducer orientations and machine parameters, posing challenges for data pooling across different raters and settings. Muscle texture analysis offers a potential means of reducing inter-rater and machine-setting variability. Scans were assessed at nine angles, controlled using a custom transducer shell and software. Scans were performed three times, and different gains were applied. All scans were performed on a muscle tissue-mimicking phantom to eliminate biological variability. Intra-angle and intra-gain variability and internal consistency were assessed via coefficient of variation (CV%) and Cronbach's alpha (αc). Spearman's (ρ) correlations were employed to determine the relationship between echogenicity and each texture feature. Entropy (angle: CV=2.7-7.6%; gain: CV=10.5%; αc=0.86), and inverse difference moment (angle: CV=3.7-9.8%; gain: CV=16.5%; αc=0.87) were less variable than echogenicity (angle: CV=6.4-19.4%; gain: CV=39.0%; αc=0.82). Angular second moment (angle: CV=17.9-116.6%; gain: CV=71.6%; αc=0.68), contrast (angle: CV=7.8-14.7%; gain: CV=41.8%;αc=0.75), and correlation (angle: CV=9.0-13.5%; gain: CV=28.6%; αc=0.49) features were generally more variable. Entropy (ρ=0.82-0.98, p≤0.011) and inverse difference moment (ρ=-0.98--0.83, p≤0.008), were more strongly correlated with echogenicity than angular second moment (ρ=-0.98--0.77, p≤0.016), contrast (ρ=0.53-0.98, p≤0.15), and correlation (ρ=-0.25--0.19, p=0.520-0.631). Entropy and inverse difference moment features may allow data sharing between laboratory and clinical settings with ultrasound machine parameters and raters of varying skill levels. Clinical and mechanistic studies are required to determine if texture features can replace echogenicity assessments.