Intraoperative measurement of cup inclination using fluoroscopy: do we need a different correction factor when using
Christian Manuel Sterneder1,2, Bernhard Springer3, Lyubomir Haralambiev4
1Adult Reconstruction and Joint Replacement Service, Hospital for Special Surgery, 535 East 70th Street, New York, NY, 10021, USA.
Introduction:
The direct anterior approach (DAA) in THA allows for intraoperative fluoroscopy imaging (IF). A previous study has shown that the intraoperative determination of radiographic cup inclination (RI) using IF during DAA THA is reliable and reproducible. However, a 5° correction factor should be applied to intraoperative measurements to adjust for parallax when using a standard 12-inch GE 9900 elite C-arm (GE Healthcare, Chicago, IL). The current study evaluates if the use of a flat panel C-arm requires a different correction factor.
Materials And Methods:
The current study included 112 patients who underwent primary THA utilizing DAA with IF for cup placement. RI was measured using intraoperative fluoroscopy images of two different C-arm models (12-inch GE 9900 elite C-arm and OEC Elite flat panel and compared to postoperative AP pelvis radiographs.
Results:
Using the 12-inch GE 9900 elite C-arm (n = 76) mean intraoperative RI measurements (35.5°, range 28-42°) were on average 6.4° lower compared to mean postoperative measurements (41.9°, range 36-49°). Using the OEC Elite flat panel C-arm (n = 36) mean intraoperative RI measurements (36.1°, range 29-40°) were on average 5.6° lower compared to mean postoperative measurements (41.7°, range 37-48°).
Conclusion:
The current study confirms that a correction factor needs to be applied to intraoperative measurements of cup inclination to adjust for parallax regardless of the type of C-arm.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Distance Corrections
Influence of Earth's Curvature and Atmospheric Refraction on Leveling
Common Leveling Mistakes and Errors