Hydroxyurea interferes negatively with triglyceride measurement by a glycerol oxidase method
Clinical Chemistry
|August 1, 1985
Summary
Hydroxyurea interferes with a specific triglyceride assay, causing falsely low results in patients with myeloproliferative diseases. Alternative triglyceride testing methods do not exhibit this interference.
Area of Science:
- Clinical Chemistry
- Biochemistry
- Pharmacology
Background:
- Hydroxyurea is a medication used to treat myeloproliferative diseases.
- Accurate triglyceride measurement is crucial for patient management.
- Certain enzymatic assays may be susceptible to interference from therapeutic agents.
Purpose of the Study:
- To investigate the cause of extremely low triglyceride concentrations observed in patients on hydroxyurea therapy.
- To identify interference in triglyceride assay methods by hydroxyurea.
Main Methods:
- Comparison of triglyceride measurements using a "cascaded" enzymatic assay (Technicon RA-1000) and a glycerol kinase-based assay (Beckman Instruments).
- In vitro addition of hydroxyurea to serum samples to assess its effect on the "cascaded" assay.
- Analysis of the enzymatic steps involved in the "cascaded" assay for potential inhibition.
Main Results:
- The "cascaded" enzymatic triglyceride assay showed falsely low results in the presence of hydroxyurea.
- Hydroxyurea demonstrated stoichiometric inhibition of glycerol oxidase within the "cascaded" assay.
- A glycerol kinase-based triglyceride assay did not exhibit interference from hydroxyurea.
Conclusions:
- Hydroxyurea interferes with the glycerol oxidase step in a specific "cascaded" enzymatic triglyceride assay.
- This interference leads to underestimation of triglyceride levels in patients receiving hydroxyurea.
- Alternative triglyceride assays, such as those using glycerol kinase, should be considered for patients on hydroxyurea therapy.