Standardized Model of Ventricular Fibrillation and Advanced Cardiac Life Support in Swine
Published on: January 30, 2020
A Comparison Between a Resuscitation Glove and Standard Manual Compressions on the Quality of Cardiovascular
Desale Tewelde Kahsay1, Miretta Tommila2, Laura-Maria Peltonen3
1Desale Tewelde Kahsay, MSc, Department of Anaesthesiology and Intensive Care, Faculty of Medicine, University of Turku, Turku, Finland.
Background:
Several audiovisual feedback (AVF) devices have been developed to monitor chest compression quality during cardiopulmonary resuscitation (CPR). However, most marketed stand-alone AVF devices are inflexible and rigid, causing discomfort and sometimes pain to the rescuers' hands.
Objective:
The objective of this study was to evaluate the effectiveness and usability of a newly developed soft and flexible resuscitation glove designed to improve the quality of chest compressions during CPR.
Methods:
We conducted a manikin-based randomized crossover study to compare the effectiveness of a newly developed AVF device (ResuGlove CPR Group) and standard CPR (Standard CPR Group) in improving the quality of chest compressions in simulated cardiac arrest cases. The usability of the newly developed ResuGlove was assessed using a System Usability Scale questionnaire.
Results:
There were no significant differences in compression depth (mean, 53.69 vs 53.28; P = .70) and compression rate (mean, 111.48 vs 113.38; P = .23) between the ResuGlove CPR and Standard CPR groups. However, the group using ResuGlove had a higher percentage of complete chest releases between compressions ( P = .008). Furthermore, the ResuGlove CPR Group had a significantly higher percentage of participants who performed chest compressions with adequate compression depth (82.8% vs 41.4%, P = .001) and compression rate (96.6% vs 72.4%, P = .012) compared with the Standard CPR Group. The ResuGlove usability score was calculated to be 70.4.
Conclusions:
The newly developed ResuGlove significantly improved the quality of certain chest compression parameters, and the device's usability score was within the acceptable range.
Related Concept Videos
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...