Study of the aberrant retention behavior of a semi-crystalline polyamide in reversed-phase liquid chromatography
Jordy D Kruijswijk1, Stefan Wijker2, Harry J A Philipsen3
1Division of Bioanalytical Chemistry, Department of Chemistry and Pharmaceutical Sciences, Amsterdam Institute of Molecular and Life Sciences, Vrije Universiteit Amsterdam, de Boelelaan 1105 1081, Amsterdam, HV, The Netherlands; Centre for Analytical Sciences Amsterdam (CASA), The Netherlands.
Abstract:
The elution behavior of semi-crystalline polyamides (PAs) in reversed-phase liquid chromatography (RPLC) was examined. RPLC of the aliphatic polyamide 4,6 (PA46) yielded a single broad peak. However, the partly aromatic polyamide X (PAX) displayed a deviating elution profile, encompassing early- and late-eluting portions, which changed in relative abundance when varying injection and gradient conditions. These bands were suspected to be due to formation of amorphous and crystalline phases, respectively. RPLC fractions of the PAs were subjected to the same RPLC system and to size-exclusion chromatography. The presumed amorphous PAX portion showed two bands in RPLC, suggesting that (largely) amorphous and crystalline phases are formed upon sample injection. Differential scanning calorimetry (DSC) demonstrated that reducing the crystallinity of PAX decreased the relative abundance of the late-eluting fraction, approaching the behavior of aliphatic polyamides. X-ray diffraction and static-light-scattering analyses confirmed the semi-crystallinity of the two solid PA samples. Although both small particles and larger aggregates were observed in solution, these findings could not be correlated to the differences in elution profile of the PAs. Cloud-point measurements indicated that the solubility of both PA46 and PAX was almost independent of temperature. Strikingly, at low column temperatures (i.e. below the depressed melting point), PAX eluted as a single broad PAX peak. To conclude, the semi-crystallinity of PAX influences its RPLC-elution behavior, and by ensuring complete dissolution of the crystalline phase useful chemical information can be extracted from the obtained chromatograms.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
13:36Simple In-House Ultra-High Performance Capillary Column Manufacturing with the FlashPack Approach
Published on: December 4, 2021
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Analyte Adsorption and Distribution
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Affinity Chromatography
High-Performance Liquid Chromatography: Elution Process
Chromatographic Methods: Terminology