Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Out-of-field neutron radiation from clinical proton, helium, carbon, and oxygen ion beams
Matteo Bolzonella1, Marco Caresana1, Andrea Cirillo1
1Department of Energy, Politecnico di Milano, Milan, Italy.
Background:
In hadron therapy, out-of-field doses, which may in the long-term cause secondary cancers, are mostly due to neutrons. Very recently, 4He and 16O beams have been added to protons and 12C ions for clinical therapy.
Purpose:
The focus of this article is to compare secondary neutron doses produced by clinical protons, 4He, 12C, and 16O ion beams.
Methods:
Ambient dose equivalent, H*(10), measurements were performed, with five types of rem counters, of the neutron field produced by the four primary ions impinging on a water phantom. This experiment was performed at the Heidelberg Ion Beam Therapy Center (HIT) in the framework of the activities of the European Radiation Dosimetry Group (EURADOS). The experimental data are normalized to both unit primary particle and target dose, and are further compared to Monte Carlo (MC) simulations performed with the FLUKA and MCNP codes.
Results:
The intensity of the neutron field increases with ion mass, and the trend is more significant in the forward direction. The minimum H*(10) for all ions, 5µSv/Gy to 10µSv/Gy, was measured in the transverse and backward directions, whereas the maximum measured value was about 1.3 mSv/Gy for primary 16O ions in the forward direction. Additional MC simulations are presented for a more detailed analysis of the rem counters' response in the presence of heavy charged fragments. In the downstream direction, for 12C and 16O ions, approximately only 30% of the instruments' counts are due to neutrons.
Conclusion:
The four extended-range instruments provide reliable and consistent results, whereas the conventional rem counter underestimates H*(10) in a neutron field with a large high-energy component. FLUKA and MCNP provide consistent predictions, within a factor of 1.6 for the downstream position and lower differences in the other cases, and are in agreement with the experimental data. It was found that under certain conditions neutrons do not represent the only secondary radiation field to be monitored; the presence of charged particles affects the performance of moderator-type neutron detectors.
More Related Videos
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
10:10Neutron Crystallography Data Collection and Processing for Modelling Hydrogen Atoms in Protein Structures
Published on: December 1, 2020
Related Concept Videos
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Biological Effects of Radiation
Isotopes and Radioisotopes
An isotope containing...
X-ray Imaging
The Electromagnetic Spectrum