Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Bioreactor Controls-I01:28

Bioreactor Controls-I

Maintaining optimal conditions within fermenters is essential for maximizing microbial productivity and ensuring process efficiency. This lesson focuses on key parameters—temperature, foam, pH, carbon dioxide, oxygen, and pressure—and their precise measurement and control strategies in fermentation systems.Temperature ControlTemperature regulation is critical due to the exothermic nature of many fermentation processes. In small laboratory fermenters, temperature is commonly monitored using...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Ultralow-Threshold Lasing in a Nanobeam Cavity Using Topological Interface State.

Nano letters·2026
Same author

Large-scale data-driven inverse EM design based on metacircuit-embedded surface.

Nature communications·2026
Same author

Identification of alpha1-oleate as a potent regulator of adipokine-dependent metabolism, in bladder cancer tissue.

Cancer & metabolism·2026
Same author

Establishment and application of electricity-based industrial supervision indicators for heavy pollution emergencies.

Journal of environmental sciences (China)·2026
Same author

Mechanical reinforcement of graphene via wrinkling.

NPJ 2D materials and applications·2026
Same author

Aging disrupts spatiotemporal coordination in the cycling murine ovary.

Nature aging·2026

Related Experiment Video

Updated: Jun 29, 2026

Improving IV Insulin Administration in a Community Hospital
12:08

Improving IV Insulin Administration in a Community Hospital

Published on: June 11, 2012

18.5K

Accuracy of a 15-day Factory-Calibrated Continuous Glucose Monitoring System With Improved Sensor Design.

Shridhara Alva1, Anuj Bhargava2, Bruce Bode3

  • 1Abbott Diabetes Care, Alameda, CA, USA.

Journal of Diabetes Science and Technology
|April 4, 2025
PubMed
Summary

The FreeStyle Libre 2 Plus and Libre 3 Plus continuous glucose monitoring systems demonstrate high accuracy over 15 days. These systems offer reliable glucose readings for adults and children, reducing vitamin C interference.

Keywords:
FreeStyle Libre 2accuracycontinuous glucose monitoringfactory calibrationoptional alarm

More Related Videos

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
08:19

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing

Published on: June 1, 2012

14.3K
Simple Continuous Glucose Monitoring in Freely Moving Mice
03:25

Simple Continuous Glucose Monitoring in Freely Moving Mice

Published on: February 24, 2023

5.0K

Related Experiment Videos

Last Updated: Jun 29, 2026

Improving IV Insulin Administration in a Community Hospital
12:08

Improving IV Insulin Administration in a Community Hospital

Published on: June 11, 2012

18.5K
Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
08:19

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing

Published on: June 1, 2012

14.3K
Simple Continuous Glucose Monitoring in Freely Moving Mice
03:25

Simple Continuous Glucose Monitoring in Freely Moving Mice

Published on: February 24, 2023

5.0K

Area of Science:

  • Medical Devices
  • Biomedical Engineering
  • Endocrinology

Background:

  • Evaluates the 15-day factory-calibrated continuous glucose monitoring (CGM) sensor in FreeStyle Libre 2 Plus and Libre 3 Plus Systems.
  • Features an improved sensor design to minimize vitamin C interference, enhancing reliability.

Purpose of the Study:

  • To assess the accuracy and performance of the FreeStyle Libre 2 Plus and Libre 3 Plus CGM systems.
  • To validate sensor performance across various age groups and glucose ranges.

Main Methods:

  • Enrolled 332 participants (2+ years) across seven US sites.
  • Collected venous and capillary blood for comparator measurements, with glycemic manipulation in older participants.
  • Calculated performance metrics including percentage of values within ±20 mg/dL or 20% and Mean Absolute Relative Difference (MARD).

Main Results:

  • Adults and pediatric participants (6-17 years) showed high accuracy: 94.2% and 94.0% within ±20 mg/dL/20% with MARDs of 8.2% and 8.1% respectively, against YSI comparator.
  • Younger children (2-5 years) achieved 86.6% within ±20 mg/dL/20% with MARD of 11.2% against SMBG.
  • Excellent performance in hypoglycemia: 97.1% (adults) and 98.0% (pediatrics) within ±15 mg/dL of YSI comparator.

Conclusions:

  • The FreeStyle Libre 2 Plus and Libre 3 Plus Systems provide accurate glucose results.
  • The 15-day wear period demonstrates reliable performance across the full glucose measurement range.