Simulated low-dose multi-detector computed tomography: spatial effects on surrogate parameters of bone strength at
Nico Sollmann1,2,3, Kai Mei4,5, Maximilian T Löffler6,7
1Department of Diagnostic and Interventional Radiology, University Hospital Ulm, Ulm, Germany. nico.sollmann@tum.de.
Abstract:
This study investigated simulated tube current reduction and sparse sampling for low-dose computed tomography (CT) regarding volumetric bone mineral density (vBMD) and cortical bone thickness (Ct.Th) of the proximal femur. Sparse sampling with dose reductions of up to 90% may still allow extraction of bone strength parameters with clinically acceptable accuracy.
Introduction:
We aimed to investigate effects of CT with simulated lowered tube current and sparse sampling on trabecular and cortical vBMD as well as Ct.Th of the entire proximal femur, its subregions, and with detailed spatial assessments.
Methods:
Clinical routine multi-detector CT (MDCT) scans covering the hips from 40 patients were used for simulations of low-dose imaging with 50% and 10% of the original tube current (D50, D10) or projections (P50, P10) combined with statistical iterative reconstruction (SIR), which were then compared against original data with full dose (D100 P100) regarding trabecular vBMD, cortical vBMD, and Ct.Th. An automated framework for multi-parametric assessments was used. Relative errors by comparing measures from original data and simulated low-dose data, regression analyses, Bland-Altman analyses, and statistical parametric mapping (SPM, to assess the spatial distribution of accuracy) were computed.
Results:
Sparse sampling enabled drastic reductions of radiation exposure (down to 10% of original imaging) while still producing determinants of bone strength with clinically acceptable relative changes. Lower biases according to Bland-Altman analyses were observed for sparse sampling compared to imaging with virtually lowered tube currents (D10 P100 versus D100 P10) regarding trabecular vBMD, cortical vBMD, as well as Ct.Th. Better accuracy across the whole proximal femur for D100 P50 than for D50 P100 and for D100 P10 than for D10 P100 was observed.
Conclusions:
Sparse sampling with SIR may enable drastic reductions of radiation exposure (up to 90% of original doses) for opportunistically measuring image-based surrogate parameters of bone strength.
More Related Videos
04:24Proper Positioning and Restraint of a Rat Hind Limb for Focused High Resolution Imaging of Bone Micro-architecture Using In Vivo Micro-computed Tomography
Published on: November 22, 2017
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography