Skin marker for MR-only surface high-dose-rate brachytherapy.
Evangelia Kaza1,2, Phillip M Devlin1,2, Desmond O'Farrell1,2
1Brigham and Women's Hospital, Dana-Farber Cancer Institute, Harvard Medical School, Boston, MA, USA.
Solid silicone rubber tubes are a viable alternative to CT markers for MRI-only surface brachytherapy. These markers are detectable and trackable on MRI and CT, enabling precise contouring for treatment planning.
Area of Science:
- Medical Imaging
- Radiation Oncology
- Biomaterials
Background:
- Surface high-dose-rate (HDR) brachytherapy requires accurate visualization of clinical targets, including human skin and fibromatosis.
- Current workflows rely on markers detectable by both MRI and CT, but existing MR markers are unsuitable for continuous surface target tracing.
- Complete MR-only workflows necessitate reliable skin markers for defining clinical target edges.
Purpose of the Study:
- To evaluate silicone rubber tubes as an alternative skin marker for MRI and CT detectability and contourability in brachytherapy.
- To assess the suitability of these markers for MR-only surface HDR brachytherapy workflows.
Main Methods:
- Silicone rubber tubes (2-3 mm diameter) were applied to phantoms, volunteers, and patients with palmar fibromatosis.
- Imaging was performed using optimized 3D PETRA and VIBE sequences with Dixon reconstruction, alongside standard CT.
- Tube detectability, conspicuity, and contouring accuracy were evaluated using a brachytherapy treatment planning system (TPS) and contouring software.
Main Results:
- Silicone rubber tubes were successfully detected on PETRA, VIBE (especially Dixon VIBE fat-only), and CT images.
- 3 mm diameter tubes demonstrated better tracking in the TPS compared to 2 mm tubes.
- MR imaging in three orthogonal planes improved tube localization within the TPS.
- Contour shapes derived from the markers generally agreed with known tube positions.
Conclusions:
- Solid silicone rubber tubes, particularly 3 mm in diameter, are suitable skin markers for MR-only surface HDR brachytherapy.
- These markers offer a viable alternative to traditional CT markers, facilitating complete MR-only treatment planning.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Imaging Studies VII: Vascular Imaging