Performance Comparison of Oximetry Devices During 6-Minute Walk Tests
Emma M Marshall1,2, Angeline Filiex3, Max Schulte2
1Ms. Marshall, Dr. Dowman, Mr. Churchward, and Dr. Khor are affiliated with Institute of Breathing and Sleep, Victoria, Australia.
None:
Background: Pulse oximetry is a common technique used to measure SpO2, with increasing numbers of portable devices of different designs and technical specifications. This study examined accuracy of commercial and medical-grade portable oximetry devices (BioBeat wristwatch, Nonin WristOx2 3150, and Heart Sure A320) in measuring SpO2 compared with the reference standard (Masimo Rad-5) during 6-min walk tests (6MWTs). Methods: This prospective study included 50 subjects, with 10 recruited for each of the five target groups, including COPD, interstitial lung disease, pulmonary hypertension, oxygen therapy users, and other cardiac/respiratory diseases. All subjects underwent 6MWT using the reference standard and the three test oximetry devices. Accuracy was measured by comparing mean nadir SpO2. Subjects were asked to rate their preference for different oximetry devices at test completion. Results: In contrast to the Masimo, the Nonin had lower nadir and mean SpO2 (P ≤ .001 for both), with higher measurements for the BioBeat (P < .001 for both) and no differences for the Heart Sure (P = .18 for nadir SpO2, P = .96 for mean SpO2). Time spent at SpO2 ≤ 88 was higher for the Nonin (70%, P < .001) but lower for the Heart Sure (58%, P < .001), compared with the Masimo (64%). There were no significant differences observed in peak heart rate measurements between the Masimo with the Nonin and BioBeat (P < .05 for both), with lower measurements using the Heart Sure (P < .001). The BioBeat was most preferable amongst subjects (n = 31), with all test oximeters having high patient satisfaction (mean \7-point Likert scale scores of 6-7). Conclusion: The Heart Sure A320 and Nonin Wrist-Ox2 3150 had comparable performance in detecting nadir SpO2 within a mean difference of 3% in comparison with the Masimo Rad-5, with potential use for remote monitoring.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...