Log file-based quality assurance method for respiratory gating system
Wonjoong Cheon1, Young Kyu Lee1, Yunji Seol2
1Department of Radiation Oncology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
Background:
As medical linear accelerator technology advances, enabling higher dose rate deliveries, hypofractionation regimens has increased. This necessitates respiratory gating systems that synchronize radiation delivery with tumor position, requiring simple rigorous quality assurance (QA) to ensure treatment accuracy and patient safety.
Purpose:
This study aimed to propose log-based QA for respiratory-gated radiation therapy using the respiratory gating system and treatment machine.
Methods:
4D CT scans were performed with a Varian motion phantom using a Varian Respiratory Gating for Scanner (RGSC). A treatment plan using 25%-75% respiratory phases with 100 MU was created and delivered to a solid water phantom. Treatment logs containing respiratory signals, beam on/off flags, and frame information were extracted from the treatment planning system's offline review. Log file analyses were conducted using in-house softwares to assess temporal synchronization between respiratory phases and beam triggers. Output measurements using a calibrated ion chamber (FC65G) were performed to evaluate dosimetric accuracy. Additionally, EPID images were acquired in cine mode and analyzed frame-by-frame to independently verify beam delivery timing.
Results:
Log file analysis revealed precise temporal synchronization, with mean time differences of 0.03 s ± 0.05 s between the planned 25% phase and beam-on, and -0.04 s ± 0.05 s between 75% phase and beam-off. The log-derived beam-on duration (2.61 s ± 0.02 s) closely matched the planned duration (2.66 s ± 0.00 s). Three-month log data showed consistent temporal accuracy, with trigger-on times remaining stable at 2.60 s ± 0.01 s across all measurements. Supporting ion chamber measurements confirmed dosimetric agreement between gating and non-gating modes (difference: 0.05 cGy ± 0.09 cGy) CONCLUSIONS: The proposed log file-based QA method demonstrated high accuracy and reproducibility in assessing respiratory gating performance. This approach provides an efficient, objective method for standardizing QA procedures in respiratory-gated radiation therapy, enhancing treatment accuracy and patient safety.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Quality Assurance
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: