Appropriate compression values for the transtibial prosthesis using the AERO prosthetic liner.
Yusuke Miyata1, Yasmin Nutchamlong1, Gary Guerra2
1Sirindhorn School of Prosthetics and Orthotics, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
A 4% compression in affordable ethylene-vinyl acetate roll-on (AERO) liners may offer comfortable transtibial prosthetic fittings. Optimal compression varies by liner thickness, with 4.95% for 3-mm and 4.5% for 5-mm liners.
Area of Science:
- Biomedical Engineering
- Prosthetics and Orthotics
Background:
- Comfortable transtibial prosthetic socket fitting relies on positive model rectification.
- Affordable ethylene-vinyl acetate roll-on (AERO) liners are used in prosthetic fittings.
Purpose of the Study:
- To evaluate if a 4% compression in 3-mm and 5-mm AERO liners provides adequate comfort and pressure distribution for users.
- To determine optimal compression values for different AERO liner thicknesses.
Main Methods:
- Utilized stump socks for volume adjustment during fitting with 4% socket compression.
- Collected pressure data using 6 force-sensing resistor (FSR400) sensors in sensitive and tolerant areas.
- Analyzed peak pressure, pressure uniformity (coefficient of variation), and socket comfort scores.
Main Results:
- Optimal compression was 4.95% for 3-mm liners and 4.5% for 5-mm liners.
- A slight similarity in pressure distribution was observed between liner thicknesses.
- The 5-mm liner potentially offered more even pressure distribution due to its thickness.
Conclusions:
- A universal 4% compression in AERO liners may be effective in clinical settings.
- Liner thickness influences pressure distribution despite identical material composition.
- Further research with more participants is needed to establish precise compression values for diverse residual limbs.
More Related Videos
08:08Oscillation and Reaction Board Techniques for Estimating Inertial Properties of a Below-knee Prosthesis
Published on: May 8, 2014
11:16Engineering Platform and Experimental Protocol for Design and Evaluation of a Neurally-controlled Powered Transfemoral Prosthesis
Published on: July 22, 2014
Related Concept Videos
Ankle Joint
Wedges
Consider using a wedge to lift a heavy slab. Here, the wedge functions by converting the applied force into a much larger force directed almost perpendicular to the initial force. This...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Angle of Twist - Elastic Range