A Blueprint for Clinical-Driven Medical Device Development: The Feverkidstool Application to Identify Children With
Evelien B van Kempen1,2, Sanne E W Vrijlandt2, Kelly van der Geest3
1Department of Paediatrics, Juliana Children's Hospital Haga Hospital, Den Haag, the Netherlands.
Abstract:
Clinical decision rules (CDRs) integrated into applications enhance diagnostic and treatment prediction support for clinicians, necessitating Confirmité Europeenne (CE)-mark certification to enter the European market. We describe the development of a CDR as a medical device, focusing on challenges from a physician's perspective exemplified by the Feverkidstool (FKT), a validated CDR for febrile children. We pursued a local process, aligned with the CE-marking process, to develop the FKT as in-house developed device. We aimed to provide a blueprint for colleagues. Medical device development, conforming the medical device regulation and performed by a multidisciplinary team, encompassed 5 stages: market scan, design, production, verification and validation and conformity assessment. Regulatory processes were continuously updated. The market scan identified a need for the FKT compared with existing applications. A prototype was designed in stage 2, further adjusted and improved based on the qualitative and quantitative results of stages 2-4. Lastly, stage 5 confirmed FKT's performance and safety. Medical device development presents challenges for physicians, requiring collaboration for technical, regulatory, and financial expertise. Multidisciplinary teamwork also poses challenges, including uncertainties regarding responsibility and timelines. After CE certification, adapting to evolving needs and ensuring data privacy highlights the ongoing nature of medical device development.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...