Validation of a Smart Ring Oximeter in Individuals With Dark Skin Pigment
Miguel Meira E Cruz1,2, Congcong Zhou3,4, Meir H Kryger5
1Sleep Unit, Centro Cardiovascular da Universidade de Lisboa, Lisbon School of Medicine, Lisbon, Portugal.
Objective:
To evaluate a ring wearable's accuracy in the measurement of oxyhemoglobin saturation measured by pulse oximeter (SpO2) compared to a simultaneous SaO2 obtained from arterial blood gases (ABGs) because oximeters may overestimate oxyhemoglobin saturation in darkly pigmented individuals resulting in occult hypoxemia.
Patients And Methods:
Circul+ (Bodimetrics), with a form factor of a ring, measures several variables for cardiorespiratory assessment (SpO2, movement, and heart rate). The design ensures that measurements are made on the palmar surface of the digit, where there is generally less pigment. Twenty-four healthy participants (8 Black, 16 non-Black) had catheters inserted into their radial arteries during the timeframe between August 2020 and January 2022. They were administered nitrogen-rich air that made them hypoxic in steps to an SaO2 of 70%. ABGs was sampled at various levels of hypoxia, and simultaneous readings were obtained from the Circul+. We compared SpO2 from Circul with measurement of ABGs by a medical grade blood gas analyzer.
Results:
There was excellent correlation between SaO2 measured by the ring oximeter and ABGs in both Black (y = 1.0174x - 1.573; R 2 = 0.9414) and Non-Black (y = 1.0209x - 2.5607; R 2 = 0.9207) participants. No differences were found in comparing the intercept and slopes of the regressions. At ABG of 70% and 100%, the SaO2 measured by the ring was calculated to be 69.6% and 100.2% for the Black participants and 68.9% and 99.5% for the non-Black participants. Bland-Altman analysis found a bias (deviation from the mean) of 0.0 for the Black participants and -0.7% for the non-Black participants.
Conclusion:
Results from this study confirmed that Circul+ oximetry accuracy seems to be independent of skin tone, perhaps because measurements are made on the palmar surface of the digit, where there is generally less pigment.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pigmentation
Melanin occurs in two primary forms: eumelanin that provides black and brown pigment and pheomelanin that provides red color. Dark-skinned individuals produce more melanin than those with pale...
Changes in Skin Color: Clinical Perspectives
Albinism
Albinism is a genetic disorder that affects (completely or partially) the coloring of skin, hair, and eyes. The defect is primarily...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...