Utility of a Card-Type Respiratory Rate Measuring Device for Spontaneously Breathing Patients
Yoshiaki Iwashita1, Shun Takeda2, Satoshi Kawashima2,3
1Department of Emergency and Critical Care Medicine, Faculty of Medicine, Shimane University, Ennyacho89-1, Izumo 693-8501, Shimane, Japan.
Abstract:
Background/Objectives: Accurate measurement of respiratory rate (RR) is critical for early detection of patient deterioration. A newly developed contactless card-type device measures RR and heart rate (HR) by detecting chest impedance changes. Although previously validated in mechanically ventilated patients with 15% RR accuracy, its performance in spontaneously breathing patients remains uncertain. Methods: This prospective observational study was conducted at the Department of Emergency and Critical Care Medicine, Shimane University Hospital, in December 2022. Patients admitted to the emergency center without invasive mechanical ventilation were enrolled. The card-type device was subsequently placed on the left chest, while the RR and HR were simultaneously recorded with an electrocardiogram monitor. Data from both devices were collected and compared. Results: Six patients were enrolled. The RR measurements from the card-type device were within a 10% difference from those measured using the standard monitor in four out of six cases. In two cases, the card-type device recorded an RR lower than that of the standard monitor, which coincided with periods of patient speaking. For HR, the card-type device was within a 10% margin of the standard monitor in two cases, but it underestimated HR in the remaining four cases, particularly during high tidal volumes or increased thoracic thickness. Conclusions: The contactless card-type device accurately measured respiratory rates within a 10% margin compared to standard monitors in most non-intubated patients, except during activities such as speaking. Further studies with larger sample sizes are warranted to confirm these findings and improve the device's performance.
More Related Videos
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Respiratory Volumes and Capacities I
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: