Noninvasive Glucose Measurements in Tissue Simulating Phantoms Using a Solid-State Near-Infrared Sensor
Ariel B Kauffman1, Ruben Shakya1, Shuai Yu1
1Department of Engineering, Rockley Photonics, Irvine, CA 92612, USA.
Sensors (Basel, Switzerland)
|April 12, 2025
Summary
This study benchmarks a new solid-state laser spectrometer for noninvasive glucose monitoring in skin. The prototype shows promising analytical utility, comparable to Fourier-transform spectrometers, despite higher spectral noise.
Area of Science:
- Biomedical Optics
- Spectroscopy
- Medical Devices
Background:
- Noninvasive glucose monitoring is crucial for diabetes management.
- Near-infrared (NIR) spectroscopy offers a promising avenue for noninvasive measurements.
- Developing robust and accurate NIR spectrometers is an ongoing challenge.
Purpose of the Study:
- To benchmark the performance of a novel solid-state laser-based NIR spectrometer.
- To compare its analytical utility against a Fourier-transform (FT) spectrometer for glucose measurements.
- To assess the feasibility of the laser platform for noninvasive human skin measurements.
Main Methods:
- Aqueous phantoms mimicking human skin composition were used for calibration and testing.
- Spectroscopic performance was evaluated by analyzing 100% lines and root mean square (RMS) noise levels.
- Partial least squares (PLS) regression was employed to assess analytical utility for glucose quantification.
Main Results:
- The solid-state laser spectrometer exhibited a median RMS noise level of 667.2 µAU, compared to 327.8 µAU for the FT spectrometer.
- A PLS model using laser data achieved a standard error of cross-validation (SECV) of 7.82 mg/dL over 1401-2238 nm.
- The laser platform's SECV was comparable to the FT spectrometer's SECV of 6.62 mg/dL (1551-2378 nm), with the difference attributed to higher spectral noise.
Conclusions:
- The solid-state laser spectrometer demonstrates significant analytical utility for noninvasive glucose monitoring.
- Its performance is competitive with established FT spectrometers, indicating potential for clinical applications.
- Further optimization to reduce spectral noise could enhance the laser platform's accuracy and selectivity.