Evaluation of an Automated Workstation for Aliquoting Whole Blood from Primary Sample Tubes
Eric Zimmerman Zuckerman1, Brian Dukek1, Dee Hunsberger1
1Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, United States.
Background:
We evaluated the Tecan Fluent Mix and Pierce Workstation as an automated platform to aliquot directly from whole blood collection tubes. The Fluent Mix and Pierce Workstation uses the Tube Rotator™, an on-deck rotisserie, to keep whole blood in suspension, and 8 Piercing Tips™ to cross the pierceable caps of collection tubes to aliquot samples.
Methods:
We performed accuracy and precision testing by aliquoting residual whole blood specimens into an on-deck gravimetric scale. To test carryover, we aliquoted alternating blanks and high calibrators for several analytes and performed mass spectrometry on the blanks. We performed 168 runs over a 21-day period to test the robustness of the platform. We aliquoted and analyzed 2 batches of samples on both our clinical platform and the Mix and Pierce Workstation to check concordance.
Results:
We observed that the Fluent Mix and Pierce Workstation can aliquot 200 µL with 0.5% inaccuracy and 0.9%CV and can aliquot 300 µL with 0.1% inaccuracy and 0.7%CV. We were unable to detect any analyte carryover using mass spectrometry. We observed no hardware failures during the instrument stress test. Clinical comparison runs indicated that the performance of aliquots made on the Fluent Mix and Pierce Workstation was equivalent to that of aliquots made by our current clinical method.
Conclusions:
Overall, our evaluation showed that the Tecan Fluent Mix and Pierce Workstation is a viable platform for aliquoting whole blood for downstream work flows and can eliminate labor-intensive sample decapping and recapping in high-throughput settings.