A wearable ring oximeter for detection of sleep disordered breathing
Miguel Meira E Cruz1, Enguo Chen2, Yong Zhou2
1Sleep Unit, Centro Cardiovascular da Universidade de Lisboa, Lisbon School of Medicine, Lisbon, Portugal and European Sleep Center, Lisbon, Portugal.
Introduction:
Wearable devices have been developed that can continuously monitor physiologic variables. One device, Circul® (Bodimetrics Corp, Los Angeles, CA), with a form factor of a ring, measures several variables (SpO2, movement, heart rate).
Objectives:
To evaluate the potential utility of this device in detecting sleep breathing disorders.
Materials And Methods:
OSA severity measured by the oxygen desaturation Index (ODI) using a wearable oximetric Circul® ring (ODI as defined by 3 % desaturation - cODI3 %) was compared with the gold standard, polysomnography (PSG) in patients suspected of having sleep-disordered breathing. The Circul data was autoscored by the device's software; a technician scored the PSG data. The sensitivity (S), and specificity (E) for the different thresholds for cODI3 % compared to the PSG AHI and PSG ODI, were calculated.
Results:
164 patients (age = 44.8 years + 12.3 (SD)) were enrolled. Using the PSG-derived AHI as the reference for classification, the best cut-off points were: OSA = AHI ≥ 5: cODI3 % ≥ 4.3 (S 87.8 %, E 93.8 %); OSA = AHI ≥ 15: cODI3 % ≥ 13.1 (S 76 %, E 100 %); OSA = AHI ≥ 30 = : cODI3 % ≥ 16.2 (S 85.7 %, E 92 %); Using the ODI from the PSG as the reference for classification, the respective cut-off points were: OSA=ODI ≥ 5: cODI3 % ≥ 4.3 (S 93.4 %, E 88.9 %); OSA=ODI ≥ 15:cODI3 % ≥ 13.1 (S85.2 %, E98.4 %); OSA=ODI ≥ 30: cODI3 % ≥ 18.7 (S 98.4 %, E 92.2 %).
Conclusions:
Circul® oximetry demonstrated good diagnostic accuracy compared to the gold standard in determining OSA severity. cODI3 % greater than 13 suggests that significant OSA might be present.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Sleep Apnea
The condition is more prevalent among...