Cardiac function evaluation in healthy volunteers and patients with implantable cardioverter-defibrillators using
Calder D Sheagren1, Naseem Shadafny2, Terenz Escartin1
1Department of Medical Biophysics, University of Toronto, 101 College St, Toronto, M5G 1L7 Ontario, Canada; Physical Sciences Platform, Sunnybrook Research Institute, 2075 Bayview Ave, Toronto, M4N 3M5 Ontario, Canada.
Background:
Implantable cardioverter-defibrillators (ICDs) cause banding artifacts around areas of B0 inhomogeneity in conventional steady-state free precession (SSFP) cine sequences. Alternatively, high-bandwidth gradient-recalled echo (GRE) cine sequences can be used to minimize artifacts in the myocardium. In this study, we assessed the bias and interobserver variability in cardiac volumes and ejection fractions between GRE cines in acquired in the presence of ICDS and ground-truth SSFP cines (without ICDs present) in a population of healthy volunteers. Further, a small cohort of ICD patients was recruited and scanned to demonstrate clinical feasibility.
Methods:
High-bandwidth GRE cine was performed in 11 healthy volunteers with taped ICDs mimicking clinical implants. After the ICD was removed, ground-truth SSFP cine was performed. Two observers separately assessed image quality metrics and contoured the cine images to return cardiac volumes and ejection fractions. Nine patients with an ICD were also scanned with the GRE cine protocol before contrast administration; data were contoured by two observers and analyzed for interobserver agreement.
Results:
In the healthy volunteer dataset, no statistically significant differences were found when comparing volumes or ejection fractions between sequences (p > 0.05). Statistically significant differences were found when comparing right ventricular ejection fraction (RVEF) (p = 0.009) and right ventricular end-systolic volume (p = 0.029) between observers, with no other significant interobserver differences. The interobserver variability of patient left ventricular ejection fraction and RVEF data was 3-4%, with lower image quality metrics for patient scans than volunteer scans.
Conclusion:
GRE cine imaging in healthy volunteers with taped ICDs demonstrated good agreement with SSFP cine, but increased interobserver variability. In patients, reducing the breath-hold duration caused a decrease in image quality, with GRE cine imaging in patients with ICDs demonstrating poorer image quality and greater interobserver variability than in healthy volunteer studies. Future work is needed to improve GRE cine image quality in patients with ICDs to reduce interobserver variability and improve clinical confidence.
More Related Videos
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...