Optimal Oversizing With the New-Generation Evolut (PRO/PRO+/FX) Self-Expanding Valves: A Multicenter Study
Silvia Mas-Peiro1, Alberto Alperi2, Ander Regueiro3,4
1Quebec Heart and Lung Institute, Laval University, Quebec City, Quebec, Canada (S.M.-P., A.P., S.M., M.A., J.R.-C.).
Higher valve oversizing in transcatheter aortic valve replacement (TAVR) with Evolut valves reduces paravalvular leaks (PVL). This study found increased oversizing correlated with lower PVL incidence, without raising complication risks.
Area of Science:
- Cardiology
- Interventional Cardiology
- Medical Devices
Background:
- Paravalvular leaks (PVLs) are a significant complication after transcatheter aortic valve replacement (TAVR).
- Valve oversizing, especially with self-expanding valves, may reduce PVL occurrence.
- The impact of oversizing with newer Evolut valve iterations on PVL is not well-established.
Purpose of the Study:
- To evaluate the effect of Evolut valve oversizing on PVL rates in a real-world setting.
- To assess the relationship between valve oversizing and post-TAVR PVL incidence.
- To determine if oversizing impacts other clinical outcomes.
Main Methods:
- A multicenter, ambispective study included patients undergoing TAVR with Evolut PRO/PRO+/FX valves.
- Oversizing degree was calculated using aortic annulus perimeter from multidetector computed tomography.
- Primary endpoint was PVL presence (mild/moderate-severe) by echocardiography; secondary endpoints included Valve Academic Research Consortium-3 defined in-hospital outcomes.
Main Results:
- 762 patients were analyzed; median oversizing was 20%.
- Higher oversizing (>20%) was associated with a lower incidence of PVL (30.2% vs. 40.9%, P=0.002).
- Increased oversizing correlated with reduced PVL risk (OR 0.95 per 1% increase, P=0.006) without affecting mortality, stroke, or permanent pacemaker rates.
Conclusions:
- Higher Evolut valve oversizing is linked to a decreased frequency of PVL after TAVR.
- This association holds true without increasing risks of other complications like permanent pacemaker implantation.
- Avoiding low oversizing degrees is recommended for Evolut valves, particularly in borderline cases.
More Related Videos
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Aortic Regurgitation III: Medical Management
Urodynamic Studies: Uroflowmetry