Paediatric high-pitch lung imaging with photon-counting detector computed tomography: a dose reduction phantom study
Michael Zellner1, Thomas Sartoretti2, Thomas Flohr2
1Department of Diagnostic Imaging, University Children's Hospital Zurich, Lenggstrasse 30, Zurich, 8008, Switzerland. michael.zellner@kispi.uzh.ch.
Photon-counting detector computed tomography (PCD-CT) allows for low-dose paediatric lung imaging. Studies show adequate image quality is maintained at 0.07 mGy using PCD-CT with specific reconstruction settings.
Area of Science:
- Medical Imaging
- Radiology
- Paediatric Imaging
Background:
- Photon-counting detector computed tomography (PCD-CT) offers potential for reduced radiation dose in pediatric lung imaging.
- Optimizing radiation dose while maintaining diagnostic image quality is crucial in pediatric CT scans.
Purpose of the Study:
- To determine the minimum radiation dose required for adequate image quality in high-pitch lung imaging using PCD-CT.
- To evaluate image quality in a pediatric chest phantom simulating a 5-year-old child.
Main Methods:
- A dual-source PCD-CT scanner was used to image a pediatric chest phantom at various radiation doses (0.01–0.45 mGy).
- Scans were performed in standard and ultra-high resolution modes with different quantum iterative reconstruction strengths.
- Image quality was assessed both subjectively by pediatric radiologists and objectively through noise and signal-to-noise ratio measurements.
Main Results:
- Adequate image quality, defined by the visibility of small peripheral structures, was maintained at a radiation dose as low as 0.07 mGy.
- No significant differences in subjective image quality were found across scan modes or reconstruction strengths.
- Ultra-high resolution mode demonstrated improved noise reduction compared to standard mode.
Conclusions:
- PCD-CT facilitates high-pitch lung imaging in pediatric patients.
- A radiation dose of 0.07 mGy, combined with quantum iterative reconstruction 4, is sufficient for adequate image quality in a 5-year-old child phantom.