Surgical ligation clip artifacts on CT scans.
Radiology
|September 1, 1985
Summary
Different surgical ligation clips were tested for computed tomography (CT) artifacts. Absorbable polydioxanone clips produced the least image degradation, making them ideal for CT imaging.
Area of Science:
- Medical Imaging
- Biomaterials Science
- Surgical Technology
Background:
- Surgical ligation clips are commonly used in medical procedures.
- The presence of metallic surgical clips can cause artifacts in medical imaging.
- Understanding the impact of different clip materials on imaging is crucial for accurate diagnosis.
Purpose of the Study:
- To evaluate and compare the computed tomography (CT) artifacts produced by surgical ligation clips made from various materials.
- To identify which clip materials result in the least image degradation during CT scans.
Main Methods:
- Surgical ligation clips constructed from tantalum, stainless steel, titanium, and polydioxanone were immersed in a water-filled phantom.
- The phantom containing the clips was scanned using computed tomography (CT).
- Image quality and artifact severity were assessed for each clip material.
Main Results:
- Tantalum clips caused severe image degradation.
- Stainless steel clips resulted in less severe degradation compared to tantalum.
- Titanium clips produced minimal artifacts.
- Polydioxanone clips, made from an absorbable polymer, were visible on CT scans with the least impact on image quality.
Conclusions:
- The material composition of surgical ligation clips significantly influences the degree of CT artifacts.
- Absorbable polydioxanone clips demonstrate superior performance with minimal artifact generation in CT imaging.
- Polydioxanone clips are a promising alternative for patients undergoing CT scans post-surgery.