Effect of Lowering Dialysate Flow Rate on Dialysis Dose and Laboratory Values
Maryn Gardner1, Nicole E Wyatt2, Neil Sanghani1
1Division of Nephrology and Hypertension, Department of Internal Medicine, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
Introduction:
The rationale behind using higher dialysate flow rates stems from the idea of maximizing a dialysis membrane's extraction ratio. Subsequent developments in dialysis membrane design, including the hollow fiber undulations, spacer yarns, and altering the fiber packing density, helped reduce blood and dialysate flow mismatch in dialysis membranes.
Methods:
This is a single-center retrospective analysis of in-center dialysis patients. Adult (age ≥18 years) in-center hemodialysis patients who were receiving dialysis at the centers 3 months prior to and 3 months following the change in dialysate flow rates were included in the study (December 2022 to May 2023). Dialysate flow rates were reduced from 800 to 600 mL/min - using the same access and dialysis membrane.
Results:
Sixty-eight patients were enrolled in the study. The change of dialysate flow rate did not result in a statistically significant change in dialysis dose (Kt/Vurea). All regression coefficients were below 0.05 (i.e., a 0.05 change in dialysis dose for flow rate 600 vs. 800 mL/min). Significant associations were found between dialysis dose and both the blood flow rate and dialysate sodium, but these were also clinically negligible differences. Analysis of patients' laboratory testing showed no statistically significant difference in serum electrolyte values (sodium, potassium, bicarbonate, blood urea nitrogen, and creatinine).
Conclusion:
In patients treated using the same dialysate filter and dialysis access, changing the dialysate flow rate from 800 to 600 mL/min does not result in clinically significant changes in either patients' measured serum electrolytes or dialysis dose.
Related Concept Videos
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Renal Drug Excretion: Effect of Urine pH, Flow Rate, and Drug pKa
The pKa of a...
Dialysis
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Factors Affecting Renal Clearance: Drug Distribution and Drug Interactions
One important factor is the relationship between renal clearance and the apparent volume of distribution. Renal clearance tends to be inversely proportional to the apparent volume of distribution. Drugs with an extensive distribution volume or those...
Heart Failure Drugs: Diuretics