Safety and Performance of OptiVantage, a CT Contrast Media Injector, in Multi-Patient Mode
Angela Tirri1, Filippo Nicolò Iannelli2, Anthony Sequeira3
1Studio Medico Associato Dott.ssa Tirri Angela sas, Licata, Italy.
Purpose:
To evaluate the safety, performance and user's satisfaction of OptiVantage®, a dual head CT Contrast Delivery System, in multi-patient mode for patients requiring a contrast-enhanced CT examination.
Patients And Methods:
A total of 100 subjects were included in this multicentre observational clinical investigation, conducted between April 20, 2023, and October 6, 2023. The primary endpoint for safety was the rate of extravasation, and the primary endpoint for performance was the success of injection assessed by the investigator. Secondary endpoints for safety were the rates of air embolism and sepsis as well as adverse events (AE) related to the injection. Other data collected included indication, set-up time, injection parameters and user's satisfaction.
Results:
The study population included 59% of women. The mean age was 63.6 ± 12.7 years (range: 18 to 83 years), with the majority of patients (55%) older than 65 years. The main indications for undergoing contrast-enhanced CT were breast cancer, colon cancer and lung cancer or nodules. The mean volume injected was 119.5 ± 14.4 mL and the injection rate ranged from 2.8 to 4.5 mL/s (mean; 3.6 ± 0.3 mL/s). No extravasation or other adverse event, including air embolism and sepsis, was reported in any of the subjects (95% CI: [0.00%, 3.62%]). All the injections (100%; 95% CI: 95.39%, 100.00%) were considered as successful for the obtention of diagnostic images. The preparation of the subject, including the setting of the patient line, took between 6 and 10 seconds in most cases (68%) and 16 to 20 seconds for 30 patients (30%). The dayset was changed for 15 subjects and in all cases, it took no more than one minute.
Conclusion:
In conclusion, this study confirms the safety and performance of the OptiVantage® Dual head in multi-use mode for contrast injection in adult patients, particularly elderly, undergoing contrast-enhanced CT.
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
09:07Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...