[Design and Verification of a Human Energy Metabolism Detection System Based on Breath-by-Breath Method]
Chendong Li1,2, Wei Fang2, Youcai Wang2
1Anhui Medical University, Hefei, 230032.
Objective:
To accurately measure human energy metabolism with high temporal resolution, a respiratory gas analysis system was designed using a breath-by-breath approach.
Methods:
Firstly, indirect calorimetry was employed in respiratory gas analysis to measure the respiratory flow and concentration signals in real-time. Secondly, oxygen consumption and carbon dioxide production were calculated through respiratory characteristic alignment and respiratory signal segmentation. Finally, metabolic indexes were calculated according to the Weir formula. Furthermore, a controlled trial was formulated for validation comparisons with the MGC ULTIMA SYSTEM PFX CARDIO2.
Results:
The results indicate that the data points of metabolic indexes measured by this system all fall within the confidence interval when compared with those of the MGC.
Conclusion:
The system has high consistency with the MGC measurement results. Thus, the system can accurately measure metabolism in real-time and provide data support for clinical nutrition assessment and therapy.