Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Harmonization of quantitative liver function evaluation using gadoxetate disodium-enhanced magnetic resonance imaging
Eriko Yoshizawa1, Akira Yamada2,3, Yukinori Okajima1
1Department of Radiology, Shinshu University School of Medicine, Matsumoto, Japan.
European Radiology
|April 17, 2025
Summary
This study developed a method to harmonize the hepatocellular uptake index (HUI) using gadoxetate disodium-enhanced MRI, ensuring consistent liver function assessment across different MRI systems. The harmonized HUI, utilizing the albumin-bilirubin linear predictor (ALBI-LP), accurately predicts liver reserve comparable to established methods.
Area of Science:
- Radiology and Imaging
- Hepatology
- Medical Physics
Background:
- Quantitative liver function assessment is crucial for clinical decision-making.
- Gadoxetate disodium-enhanced MRI offers a non-invasive method to evaluate liver function via the hepatocellular uptake index (HUI).
- Variability in MR systems hinders the comparability of HUI measurements, necessitating harmonization.
Purpose of the Study:
- To develop and validate a clinically applicable harmonization method for the hepatocellular uptake index (HUI) in gadoxetate disodium-enhanced MRI.
- To ensure consistent and comparable quantitative liver function assessment across diverse MR imaging systems.
- To establish a harmonization technique that correlates HUI with established measures of liver reserve, such as indocyanine green disappearance rate (ICG-PDR).
Main Methods:
- A retrospective study included 498 patients undergoing EOB-MRI, HUI, ICG-PDR, and albumin-bilirubin linear predictor (ALBI-LP) measurements on six different MR systems.
- A harmonization method was developed using ALBI-LP to estimate liver reserves corresponding to ICG-PDR.
- Statistical analysis of residuals was employed to validate the harmonization method and derive conversion factors for HUI.
Main Results:
- A statistically significant linear correlation was found between HUI, ICG-PDR, and ALBI-LP across all MR systems.
- A harmonizing equation (h-HUI = HUI・(Slope2'/-1.425)・0.955) was derived, where Slope2' is the regression slope between HUI and ALBI-LP for each system.
- Harmonized HUI (h-HUI) using ALBI-LP demonstrated a significantly smaller estimation error for ICG-PDR compared to non-harmonized HUI, achieving comparability with h-HUI derived from ICG-PDR.
Conclusions:
- The ALBI-LP-based harmonization of HUI provides a clinically applicable method for standardizing quantitative liver reserve prediction using EOB-MRI.
- This harmonization ensures the comparability of MR devices, enabling reliable liver function assessment across different imaging platforms.
- The developed method allows for accurate estimation of quantitative liver function, equivalent to the indocyanine green clearance test, using harmonized HUI across diverse MR systems.
Keywords:
Contrast mediaGadoxetic acid disodiumIndocyanine greenLiver function testMagnetic resonance imaging