Related Experiment Video
Updated: Jul 16, 2026

Terahertz Microfluidic Sensing Using a Parallel-plate Waveguide Sensor
Published on: August 30, 2012
Coupled split square resonator based metamaterial sensor for realtime monitoring of air bubbles in IV channels
Musa Ahmad1, Mohammad Tariqul Islam2, Touhidul Alam3,4
1Pusat Sains Angkasa, Institut Perubahan Iklim, Universiti Kebangsaan Malaysia (UKM), 43600, Bangi, Selangor, Malaysia.
Abstract:
Air embolism caused by undetected air bubbles in intravenous (IV) lines poses a critical risk in medical treatments, necessitating accurate and reliable detection systems. This study addresses this challenge by developing a novel metamaterial-based sensor for real-time monitoring of air bubbles in IV channels. The sensor employs a Single Negative Complementary Split Square Resonator (CSSR) coupled with a polymer-based microfluidic channel (PMC) to detect dielectric changes caused by air bubbles in flowing saline. Unlike traditional optical or ultrasonic methods, the proposed sensor offers a compact, cost-effective, and fabrication-friendly solution while maintaining high sensitivity of 78% and precision. Experimental validation demonstrated its capability to detect even small air bubbles, with resonant frequency shifts observed between 6.6 GHz and 6.9 GHz. This innovation provides a robust tool for enhancing patient safety by preventing air embolisms, with potential applications in broader biomedical sensing scenarios. By addressing a critical medical need through advanced metamaterial technology, this work contributes significantly to both healthcare monitoring and sensor design.