Bacterial contamination of ultrasound probes in emergency departments: a multi-centre observational study
Takahiro Yamanaka1, Ryo Yamamoto1, Keitaro Yajima1
1Department of Emergency and Critical Care Medicine, Keio University School of Medicine, Tokyo, Japan.
Background:
Ultrasound is frequently used in emergency departments (EDs) for patient evaluation and diagnosis. Despite the risk of probe contamination from body fluids and blood, the rate of such contamination remains unclear.
Aim:
This study aimed to evaluate bacterial contamination of ultrasound probes in EDs, focusing on hospital types and reprocessing methods.
Methods:
A multi-centre prospective observational study was conducted at a university hospital, a non-academic tertiary hospital, and a regional hospital in 2023. Samples were collected from probes used on ED patients. Reprocessing methods included water-moistened wipes alone, water-moistened wipes with ethanol wipes, quaternary ammonium wipes alone, and quaternary ammonium wipes with ethanol or hypochlorite wipes. Outcomes included the level of bacterial contamination, measured by colony-forming units (cfu) per total surface area of each probe, and resistant bacterial strains.
Findings:
The median cfu was 10 (IQR: 0-50) at the university hospital, 40 (10-135) at the non-academic tertiary hospital, and 30 (1-95) at the regional hospital. By reprocessing method, the median cfu was 20 (1-90) for water-moistened wipes alone, 10 (0-20) for water-moistened wipes and additional ethanol wipe, 90 (40-180) for quaternary ammonium wipes alone, and 20 (1-50) for quaternary ammonium wipe and additional ethanol or hypochlorite wipe. Resistant bacterial strains were found on 18.2% of probes.
Conclusion:
High levels of bacterial contamination, including resistant strains, were observed on ultrasound probes in EDs, regardless of facility type and reprocessing method.
More Related Videos
10:38Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
Published on: January 16, 2019
05:49Tracking Microbial Contamination in Retail Environments Using Fluorescent Powder - A Retail Delicatessen Environment Example
Published on: March 5, 2014
Related Concept Videos
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Bacterial Signaling
Transmission-based Precautions I: Contact, Enteric, and Droplets
Contact Precautions:
Contact precautions are the measures taken to prevent the transmission of infectious agents, especially epidemiologically important microorganisms such as MRSA or influenza, primarily transmitted through direct or indirect contact with an...
Standard Precaution
Hand hygiene is the most crucial means to prevent the transmission of disease. Employers are legally required to provide their workers with personal protective equipment (PPE) to minimize exposure or contact with...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Ultrasonography
During an ultrasonography procedure, a handheld device called...