Dynamic Navigation for Dental Implant Placement
Published on: September 13, 2022
Accuracy of Orthodontic Miniscrew Implantation Assisted by Dynamic Navigation Technology Combined With Cone Beam
Xinze Zhang1, Yingying Su1, Ruiqing Wu1
1Dentistry, Beijing Tiantan Hospital, Capital Medical University, Beijing, CHN.
Objective:
To evaluate the accuracy of orthodontic miniscrew placement using a dynamic navigation system combined with cone beam computed tomography (CBCT) imaging in porcine mandibles.
Methods:
The Porcine mandible was scanned using CBCT. Virtual implant positions were designed using the planning software based on CBCT data. A total of 30 orthodontic miniscrews were placed in the porcine mandibles with the aid of the dynamic navigation system. A CBCT scan was taken after implantation. The deviations at entry points, endpoints. and angle deviation between the actual and planned miniscrew locations were measured by comparing the pre- and post-operation CBCT.
Results:
CBCT images of post-surgical dynamic navigation guided-miniscrew placement showed no root contact (0/30). The average linear deviations at the miniscrew entry points and endpoints were 1.8 ± 0.7 mm and 1.8 ± 0.6 mm, respectively. The average angular deviation was 4.1° ± 1.6°.
Conclusion:
The dynamic navigation system achieves acceptable accuracy for orthodontic miniscrew placement in porcine mandibles, reducing risks associated with implantation.