Commissioning of a Commercial Secondary Dose Check Software and Clinical Implementation for the Magnetic
José Alejandro Rojas-López1,2, Alexis Cabrera-Santiago3,4, Jorge Ramiro Corral-Beltrán5
1Department of Radiotherapy, Hospital Angeles Puebla, Av. Kepler, Reserva Territorial Atlixcáyotl, Puebla, México.
Journal of Medical Physics
|April 21, 2025
Summary
This study commissioned ThinkQA (TQA) software for magnetic resonance-guided linear accelerator (MR-linac) secondary dose calculations. TQA was clinically implemented for adaptive radiotherapy using a 3%/2 mm gamma criterion, ensuring treatment quality with high-to-low-density interfaces.
Area of Science:
- Medical Physics
- Radiation Oncology
- Medical Imaging
Background:
- Magnetic resonance-guided linear accelerators (MR-linacs) offer advanced precision in radiation therapy.
- Accurate secondary dose calculation software is crucial for ensuring treatment efficacy and patient safety with MR-linacs.
- Commissioning new software requires rigorous validation against established guidelines and clinical data.
Purpose of the Study:
- To commission the ThinkQA (TQA) software as a secondary dose calculation engine for an MR-linac.
- To validate TQA's performance against Medical Physics Practice Guideline 5.a. (MPPG5a) and clinical measurements.
- To establish appropriate gamma criteria for clinical implementation in adaptive radiotherapy workflows.
Main Methods:
- MPPG5a tests, including dose in inhomogeneities, beam profiles, and depth dose curves, were performed and compared between Monaco and TQA.
- Five intensity-modulated radiotherapy (IMRT) plans were evaluated using varying gamma criteria.
- Clinical data from 17 patients and measurements from five patients using ArcCheck were analyzed and compared with TQA calculations.
Main Results:
- Most MPPG5a tests met tolerances, with exceptions in dose profiles and fields with extensive blocking.
- A 3%/2 mm gamma criterion was found suitable for IMRT plans with high-to-low-density interfaces, unlike tighter criteria.
- Clinical case calculations showed high agreement between TQA and ArcCheck, with minor discrepancies in a specific prostate case.
Conclusions:
- ThinkQA (TQA) is successfully commissioned as a secondary dose calculation software for MR-linac.
- TQA is clinically implemented for adaptive treatment workflows using a 3%/2 mm gamma criterion (95% tolerance, 90% action limit).
- The chosen criteria maintain treatment quality, even with complex density interfaces.