Related Experiment Video
Updated: May 10, 2025

Quantitative Analysis of Vacuum Induction Melting by Laser-induced Breakdown Spectroscopy
Published on: June 10, 2019
Laser-spectroscopy testbed for impurity monitoring in liquid metal-cooled fast reactors
A Burak1, L Frigerio1,2, J Brown1
1Department of Nuclear Engineering and Radiological Sciences, University of Michigan, 2355 Bonisteel Boulevard, Ann Arbor, Michigan 48109, USA.
Abstract:
A significant challenge in sodium-cooled fast reactors is controlling impurities, in particular oxygen impurities, within the sodium coolant, as they can accelerate corrosion and indicate leaks. Optical methods offer the potential to rapidly detect small concentrations of both gaseous and metal impurities that accelerate corrosion, plug coolant channels, and lead to increased activation of isotopes in the coolant. We present the design and performance of an apparatus designed to enable the application of multiple optical analytical techniques, such as laser-induced breakdown spectroscopy, to detect elemental impurities in the sodium melt with high sensitivity. We experimentally demonstrate the detection of characteristic sodium and oxygen spectral lines in liquid sodium, which sets the stage for the optimization of its analytical sensitivity. A robust sensor of this type integrated with the sodium cooling loop has the potential to significantly improve the safety and operational efficiency of generation IV nuclear reactors.