Related Experiment Video
Updated: May 10, 2025

Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
Nonconcurrent Control Use in FDA Approval of High-Risk Medical Devices
Maryam Mooghali1, Sanket S Dhruva2,3, Hollin R L Hakimian4
1Department of Internal Medicine, Yale School of Medicine, New Haven, Connecticut.
Importance:
The US Food and Drug Administration (FDA) may consider historical controls, objective performance criteria, or performance goals as alternatives to concurrent control groups in clinical studies evaluating safety and/or effectiveness of high-risk medical devices. However, nonconcurrent control use has limitations.
Objective:
To examine frequency, justification, and clinically relevant details for use of nonconcurrent controls in pivotal studies supporting FDA approval of high-risk medical devices.
Design, Setting, And Participants:
Cross-sectional study from 2019 through 2023 of original high-risk therapeutic medical device approvals in the US.
Main Outcomes And Measures:
Frequency, justification, and clinically relevant details for use of historical controls, objective performance criteria, and performance goals and whether they had been established by the FDA.
Results:
From 2019 to 2023, the FDA approved 101 original high-risk therapeutic medical devices, of which 13 (12.9%) were not approved on the basis of any pivotal study. Among the remaining 88 approvals, 33 (37.5%) were based on pivotal studies that used concurrent controls for all analyses, 3 (3.4%) on single-group studies without the use of nonconcurrent controls, and 52 (59.1%) were based on at least 1 analysis using historical controls, objective performance criteria, or performance goals for primary safety and/or effectiveness end points. These 52 premarket approvals included 79 primary safety or effectiveness analyses using nonconcurrent controls: 9 historical controls (11.4%), 7 objective performance criteria (8.9%), and 63 performance goals (79.7%). Overall, 3 analyses (3.8%) provided justification for use of nonconcurrent controls. Clinically relevant details were provided for 19 analyses (24.1%) with respect to historical data publication dates and for 10 (12.7%) with respect to comparison of patient characteristics between pivotal studies and nonconcurrent controls. Four of 7 objective performance criteria (57.1%) and 0 of 63 performance goals had been established by the FDA.
Conclusions And Relevance:
This cross-sectional study of high-risk therapeutic medical device approvals by the FDA over the past 5 years found that most approvals were based on analyses using nonconcurrent controls with limited justification and clinically relevant details, potentially creating uncertainty for patients, clinicians, and payers regarding the evidence establishing device safety and effectiveness.
Related Concept Videos
Drug Regulation
Clinical Trials: Overview
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Drug Control Governance: Regulatory Bodies and Their Impact
Pharmacovigilance
This process, termed pharmacovigilance, aims to detect, evaluate, and minimize harmful effects related to medication use. The data collection for pharmacovigilance depends on spontaneous reporting systems, where healthcare professionals or patients voluntarily report suspected ADRs.
In some cases, there...