Biomechanical comparison of plate, intramedullary screw, and headless compression screw fixation in Danis-Weber type
Chen-Yuan Yang1,2, Wei-Chun Hong1,2, Ming-Tzu Tsai3
1Department of Orthopedic Surgery, Kuang Tien General Hospital, Taichung, 433, Taiwan.
Background:
Weber type B transsyndesmotic fractures are the most common type of lateral malleolar fractures, a common ankle fracture type. Although several surgical techniques are available for fixation, no clinical consensus exists regarding the optimal fixation method.
Objective:
This study evaluated the biomechanical differences in fixation stability among one-third tubular plates, 3.5-mm intramedullary screws, and 4.5-mm intramedullary headless compression screws for Weber type B lateral malleolar fractures using an in Vitro mechanical analysis.
Materials And Methods:
In total, 24 artificial fibula bone specimens were used. The specimens were subjected to Weber type B lateral malleolar fractures using a saw blade and randomly divided into three groups treated with different fixation methods, with eight specimens per group: (1) One-third tubular plate, (2) 3.5-mm intramedullary screw, and (3) 4.5-mm intramedullary headless compression screw. The specimens were subjected to bending and torsional mechanical testing using a materials testing system. The bending test was evaluated using 10-mm displacement force (Newtons) and bending stiffness (Newtons/millimeter) as the assessment parameters. For the torsional test, 20° rotation torque (Newtons/millimeter) and torsional stiffness (Newton∙millimeters/degree) were used as evaluation parameters. The experimental results were analyzed using the Kruskal-Wallis test. When significant differences were found, post hoc pairwise comparisons were conducted using the Dunn-Bonferroni approach.
Results:
In the bending test, no statistical differences were observed between the three fixation methods, whether evaluated by 10-mm displacement force or bending stiffness. Nevertheless, for the 20° rotation torque assessment in the torsional test, both the one-third tubular plate (1360.31 ± 221.56 N·mm, median ± interquartile range) and the 4.5-mm intramedullary headless compression screw (1420.41 ± 281.95 N·mm) exhibited similar resistance, considerably outperforming the 3.5-mm intramedullary screw (605.80 ± 165.11 N·mm) in fracture fixation strength. In the torsional stiffness assessment, both the one-third tubular plate (67.67 ± 15.39 N·mm, median ± interquartile range) and the 4.5-mm intramedullary headless compression screw (62.44 ± 17.36 N·mm) exhibited equivalent resistance and were substantially superior to the 3.5-mm intramedullary screw (25.90 ± 5.1 N·mm) in their ability to fix fractures.
Conclusion:
The 4.5-mm intramedullary headless compression screw provides strong mechanical stability and a lower risk of soft tissue complications in clinical practice. Therefore, this headless compression screw can be recommended as a fixation option for Weber type B lateral malleolar fractures.